An extended bore length solid-state digital-BGO PET/CT system: design, preliminary experience, and performance
John A Kennedy1,2, Tala Palchan-Hazan3, Quentin Maronnier4
1Department of Nuclear Medicine, Rambam Health Care Campus, P.O.B. 9602, 3109601, Haifa, Israel. j_kennedy@rambam.health.gov.il.
This study introduces a new digital-Bismuth Germanium Oxide (BGO) PET/CT system with a 32 cm axial field-of-view, demonstrating high sensitivity and spatial resolution. The system enables rapid clinical imaging using low-dose radiotracers.
Area of Science:
- Medical Imaging
- Nuclear Medicine
- Positron Emission Tomography (PET)
Background:
- Advancements in PET/CT technology are crucial for improving diagnostic accuracy and patient outcomes.
- Solid-state detectors offer potential advantages in sensitivity and spatial resolution compared to traditional systems.
Purpose of the Study:
- To evaluate the performance characteristics of a novel solid-state PET/CT system utilizing Bismuth Germanium Oxide (BGO) scintillating crystals.
- To assess the system's capabilities over an extended 32 cm axial field-of-view (FOV) for high spatial resolution and sensitivity.
Main Methods:
- Performance was evaluated against NEMA NU 2-2018 and EARL Standards, including spatial resolution, scatter fraction (SF), noise equivalent count rate (NECR), and image quality (IQ).
- Bayesian penalized-likelihood reconstruction (BPL) was employed for IQ assessment, with EARL Standards 2 studies utilizing BPL and a deep learning algorithm.
- Patient scans with low-dose radiotracers ([18F]FDG, [68Ga]Ga-DOTA-TATE, [82Rb]RbCl) were acquired with total scan times of ≤ 8 minutes.
Main Results:
- The system achieved a NEMA sensitivity of 47.6 cps/kBq and spatial resolution ≤3.8 mm (OSEM).
- Peak NECR was 485.2 kcps, with SF at 35.6% and count rate accuracy of 2.16%.
- Good image quality was observed in patient scans, with EARL Standards 2 criteria met and Esser phantom features ≥4.8 mm resolved within 2 minutes per bed position.
Conclusions:
- The digital-BGO PET/CT system offers superior NEMA sensitivity and high count rates within its class.
- Its performance enables fast clinical acquisitions with reduced radiotracer doses, showcasing its clinical utility.