Purpose:
This paper presents a novel computational approach to optimize gradient array performance for a given pulse sequence. Specifically, we propose an electromagnetic (EM) approach that minimizes eddy losses within the cryostat while maintaining key performance parameters such as field linearity, gradient strength, and imaging region's dimension and position.
Methods:
High-resolution EM simulations on the cryostat's surface are deployed to compute the net EM fields generated by each element of a gradient array coil at different frequencies. The computed fields are stored and combined for each frequency to form a quadratic vector-matrix-vector computation. The overall time-average eddy power loss within the cryostat assembly for arbitrary pulse sequences is computed using frequency domain superposition.
Results:
The proposed approach estimates and regulates eddy power losses within the cryostat assembly. When compared to the stray field minimization approach, it can achieve over twice the reduction in eddy power loss. The proposed approach eliminates the need to sample the stray fields on the cryostat surface, which the number and position of the samples would be challenging when designing tunable array coils with capabilities that disrupt field symmetries. Additionally, the loss calculation considers the entire cryostat assembly rather than just the inner cylindrical surface of the warm shield.
Conclusion:
Our findings highlight the efficacy of an on-the-fly tuning method for the development of high-performance whole-body gradient array coils, effectively mitigating eddy losses within the cryostat and minimizing stray fields outside the coil assembly. This approach proves particularly advantageous for array coils with variable feeding currents.
Related Concept Videos
Magnetic Damping
If, however, the bob is a slotted metal plate, the magnet produces a much smaller effect. When a slotted metal plate enters the field, an emf is induced by the change in flux; however, it is less effective because the slots limit the...
Eddy Currents
Other major applications of eddy currents appear in metal detectors and the braking systems of trains and roller...
Magnetic Field Due To A Thin Straight Wire
Magnetic Field Of A Current Loop
Force On A Current Loop In A Magnetic Field
Magnetic Field of a Solenoid
Consider a solenoid with 100 turns wrapped around a cylinder of...