Related Experiment Video
Updated: Jul 18, 2026

An Anesthesia, Surgery, and Harvest Method for the Evaluation of Transpedicular Screws Using an In Vivo Porcine Lumbar Spine Model
Published on: May 31, 2017
Biomechanical evaluation of a novel atlas polyaxial transverse connecting screw system, an in vitro human cadaveric
Yun-Lin Chen1, Wen-Jie Lu2, Yu-Li Ma3
1Department of Spine Surgery, Ningbo No.6 Hospital, Ningbo, China.
Purpose:
To introduce a novel transverse connecting screw system, and to evaluate the biomechanical stability of the novel screw system using human cadaveric specimens.
Methods:
Six fresh-frozen cadaveric upper cervical spines were used in our study. Every specimen was tested under 5 conditions: intact group; unstable group; C1 to C2 screw rod system group; C1 to C2 + crosslink system group; atlas polyaxial transverse connecting screw (APTCS) system.
Results:
Compared with the intact state, C1 to C2 screw rod system, C1 to C2 + CL system and APTCS showed statistically decrease range of motion in all directions except for the unstable group under posterior extension direction (P < .05). APTCS group has the least range of motion in all directions (P < .001).
Conclusion:
The APTCS system was able to restore stability to the atlantoaxial joint. APTCS system has the advantages of easy installation, convenient bone grafting, and strong biomechanical strength.