Impact of a cuff-based device calibration method on the agreement between invasive and noninvasive aortic and
Daniel Bia1, Federico Salazar2, Luis Cinca2
1Departamento de Fisiología, Centro Universitario de Investigación, Innovación y Diagnóstico Arterial (CUiiDARTE), Facultad de Medicina, Universidad de la República, Montevideo, Uruguay.
Introduction:
Brachial cuff-based methods are increasingly used to estimate aortic systolic blood pressure (aoSBP). However, there are several unresolved issues.
Aims:
to determine to what extent the scheme used to calibrate brachial records (1) can affect noninvasive obtained aoSBP levels, and consequently, the level of agreement with the aoSBP recorded invasively, and (2) how different ways of calibrating ultimately impact the relationship between aoSBP and cardiac properties.
Methods:
brachial and aortic blood pressure (BP) was simultaneously obtained by invasive (catheterisation) and noninvasive (brachial oscillometric-device) methods (89 subjects). aoSBP was noninvasive obtained using three calibration schemes: 'SD': diastolic and systolic brachial BP, 'C': diastolic and calculated brachial mean BP (bMBP), 'Osc': diastolic and oscillometry-derived bMBP. Agreement between invasive and noninvasive aoSBP, and associations between BP and echocardiographic-derived parameters were analysed.
Conclusions:
'C' and 'SD' schemes generated aoSBP levels lower than those recorded invasively (mean errors: 6.9 and 10.1 mmHg); the opposite was found when considering 'Osc'(mean error: -11.4 mmHg). As individuals had higher invasive aoSBP, the three calibration schemes increasingly underestimated aoSBP levels; and viceversa. The 'range' of invasive aoSBP in which the calibration schemes reach the lowest error level (-5-5 mmHg) is different: 'C': 103-131 mmHg; 'Osc': 159-201 mmHg; 'SD':101-124 mmHg. The calibration methods allowed reaching levels of association between aoSBP and cardiac characteristics, somewhat lower, but very similar to those obtained when considering invasive aoSBP. There is no evidence of a clear superiority of one calibration method over another when considering the association between aoSBP and cardiac characteristics.
Related Concept Videos
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Measurement of Blood Pressure
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...