Related Experiment Video
Updated: Jul 9, 2025

07:12
Development of a Gaze-Contingent Display Framework Designed for Perceptual and Oculomotor Research with Simulated Central Vision Loss
Published on: April 11, 2025
379
Remote Monitoring of Visual Function in Patients with Maculopathy: The Aphelion Study.
Anthony Joseph1, Mark Bullimore2, Faye Drawnel3
1Retina Associates, Tucson, AZ, USA.
Ophthalmology and Therapy
|November 28, 2023
Summary
The myVisionTrack expanded version (mVTx) tests for visual acuity (VA) and contrast sensitivity are accurate and repeatable for remote patient monitoring. These self-testing tools show potential for detecting maculopathy progression between clinic visits.
Area of Science:
- Ophthalmology
- Medical Technology
- Digital Health
Background:
- Remote monitoring of vision aids in detecting disease activity in maculopathy patients.
- The shape discrimination hyperacuity (SDH) test is a key tool for remote vision assessment.
- Evaluating novel self-testing tools is crucial for expanding remote patient monitoring capabilities.
Purpose of the Study:
- To determine the in-clinic accuracy and repeatability of three myVisionTrack expanded version (mVTx) self-testing devices.
- To assess the usability and performance of mVTx tests for visual acuity (VA) and contrast sensitivity.
- To compare the mVTx tests against standard clinical vision tests.
Main Methods:
- A prospective, single-arm study (Aphelion) included 122 adults with maculopathy at two US sites.
- Participants performed mVTx tests (tumbling E, Landolt C, contrast sensitivity, SDH) and standard clinical tests.
- Bland-Altman analyses assessed test-retest repeatability and agreement between mVTx and clinical tests; a usability survey was also administered.
Main Results:
- The mVTx tests demonstrated good repeatability, with 95% limits of agreement (LoA) ranging from ±0.18 to ±0.32 logMAR/logCT.
- Agreement with standard clinical tests was acceptable, with LoA for VA tests between ±0.35 to ±0.39 logMAR and contrast sensitivity at ±0.30 logCT.
- Most participants (85%) found the tests easy to learn, with the tumbling E test being the most user-friendly.
Conclusions:
- The mVTx visual acuity tests are accurate and repeatable, validating their use in remote monitoring.
- These self-testing tools show promise for detecting maculopathy progression between clinical appointments.
- The mVTx tests, alongside the SDH test, can enhance remote patient management for individuals with maculopathy.