Ambulatory Long-Term Oxygen Therapy in Patients with Severe COPD: A Randomized Crossover Trial to Compare
Daniel Sebastian Majorski1, Saba Gul Khan1, Sarah Bettina Stanzel1
1Department of Pneumology, Cologne Merheim Hospital, Kliniken der Stadt Köln gGmbH, Witten/Herdecke University, Cologne, Germany.
Background And Methods:
Constant-minute-volume and constant-bolus devices serve as two different means of portable oxygen conservation. A prospective randomised crossover study was conducted in COPD GOLD IV patients to investigate the effect of these two devices on dyspnea, oxygenation and 6-minute walking test (6MWT) distance. The primary endpoint was the final operating level required (operating level range 1-5 for both devices) by either device to meet the success criteria for mobile oxygen therapy, as outlined in the British Thoracic Society guidelines (SpO2 ≥90% throughout 6MWT; ≥10% increase in walking distance from baseline; improvement in BORG of at least 1 point from baseline).
Results:
Twenty-five patients were enrolled in the study and randomly assigned to one of two sequences involving the use of each type of portable oxygen conservation device. 14 female, 67.9 years (±7.8); FEV1: 27.3%pred. (±8.4); PaO2 at rest without oxygen: 50.3mmHg (±5.9). For both systems, 24/25 patients (96%) were successfully recruited. The mean operating-level difference when success criteria were met was -0.58 in favor of the constant bolus device (95% CI: -0.88 to -0.28, P <0.001). Secondary endpoints (walking distance, respiratory rate and BORG dyspnea) showed no statistically significant or clinically relevant differences. An algorithm created especially for this study showed a high success rate in terms of titration for the required operating level.
Conclusion:
Both portable oxygen-conserving devices met the success criteria in 96% of patients in the 6MWT when they were titrated to the correct level. The constant-bolus device required a significantly lower operating level to achieve the success criteria, hereby reducing energy consumption. Individual titration of the respective device is recommended, which can be facilitated by the novel titration algorithm described here.
More Related Videos
00:07Home-Based Prescribed Pulmonary Exercise in Patients with Stable Chronic Obstructive Pulmonary Disease
Published on: August 24, 2019
07:09Assessment of Pulmonary Capillary Blood Volume, Membrane Diffusing Capacity, and Intrapulmonary Arteriovenous Anastomoses During Exercise
Published on: February 20, 2017
Related Concept Videos
COPD: Management Using Bronchodilators and Corticosteroids
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Chronic Obstructive Pulmonary Disease-IV: Assessement and Diagnostic Studies
Medical History
One-Compartment Open Model for IV Bolus Administration: Estimation of Clearance
In the one-compartment open model for intravenous (IV) bolus administration, clearance is estimated by dividing the elimination rate by the plasma drug concentration. This equation leverages the elimination rate constant and the apparent...
Treatment for Pulmonary Arterial Hypertension: Oxygen Therapy for Respiratory Failure
Oxygen therapy is vital in increasing and maintaining blood oxygen levels in PAH patients. As a result, it aids in reducing fatigue,...
One-Compartment Open Model for IV Bolus Administration: Estimation of Elimination Rate Constant, Half-Life and Volume of Distribution