Assessing Agreement and Variability Among Alternative Devices for Intraocular Pressure Measurement: A Comparative
Chen-Wei Lin1,2, Jin-Jhe Wang2, Chien-Hsiung Lai2,3,4,5
1Department of Medical Education, Chang Gung Memorial Hospital, Chiayi, 61363, Taiwan.
Purpose:
Intraocular pressure (IOP) measurement is critical in diagnosing and managing eye conditions. This study aims to assess the comparability of three alternative devices for measuring IOP: Noncontact tonometer, Icare rebound tonometer, and Tono-Pen.
Patients And Methods:
A cross-sectional study included 172 adult participants (87 males and 85 females) who underwent IOP and central corneal thickness (CCT) assessments. IOP was measured using Noncontact (Canon TX-20), Icare (Icare TA01i), and Tono-Pen (Tonopen XL). CCT was measured with the built-in pachymetry of the Noncontact tonometer. Correlation coefficients and Bland-Altman analyses were conducted to assess the relationships and agreements between these tonometers. Participants were grouped based on IOP and CCT levels. The mean of the standard deviation of the three tonometer results was calculated to evaluate measurement result variability. One-way analysis of variance was conducted for comparing between the groups.
Results:
IOP measurements among the three devices were not significantly different, indicating their comparability. Correlation analysis revealed strong correlations between the tonometers. Bland-Altman analysis showed good agreement, with the Icare rebound tonometer and Tono-Pen exhibiting narrower limits of agreement. Furthermore, IOP levels influenced measurement result variability, with higher IOP levels associated with greater variance.
Conclusion:
This study demonstrates that the alternative devices examined can provide reliable IOP measurements. It highlights the potential of these alternative devices for IOP measurement. These findings have implications for clinical practice, offering practitioners additional tools for accurate IOP assessment.
Related Concept Videos
Assessment of blood pressure in brachial artery(two-step method)
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...