Photon-counting detector CT allows significant reduction in radiation dose while maintaining image quality and noise
Achala Donuru1, Tetsuro Araki1, Farouk Dako1
1Department of Radiology, Hospital of the University of Pennsylvania, Philadelphia, PA 19104, USA.
Purpose:
To investigate if clinical non-contrast chest CT studies obtained with PCD CT using much lower radiation exposure can achieve the same image quality as with the currently established EID protocol.
Materials/Methods:
A total of seventy-one patients were identified who had a non-contrast chest computed tomography (CT) done on PCD CT and EID CT scanners within a 4-month interval. Five fellowship trained chest radiologists, blinded to the scanner details were asked to review the cases side-by-side and record their preference for images from either the photon-counting-detector (PCD) CT or the energy-integrating detector (EID) CT scanner.
Results:
The median CTDIvol for PCD-CT system was 4.710 mGy and EID system was 7.80 mGy (p < 0.001). The median DLP with the PCD-CT was 182.0 mGy.cm and EID system was 262.60 mGy.cm (p < 0.001). The contrast to noise ratio (CNR) was superior on the PCD-CT system 59.2 compared to the EID-CT 53.3; (p < 0.001). Kappa-statistic showed that there was poor agreement between the readers over the image quality from the PCD and EID scanners (κ = 0.19; 95 % CI: 0.12 - 0.27; p < 0.001). Chi-square analysis revealed that 3 out of 5 readers showed a significant preference for images from the PCDCT (p ≤ 0.012). There was no significant difference in the preferences of two readers between EID-CT and PCD-CT images.
Conclusion:
The first clinical PCD-CT system allows a significant reduction in radiation exposure while maintaining image quality and image noise using a standardized non-contrast chest CT protocol.
More Related Videos
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
10:24Neutron Radiography and Computed Tomography of Biological Systems at the Oak Ridge National Laboratory's High Flux Isotope Reactor
Published on: May 7, 2021
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation I: X-ray and CT
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
X-ray Imaging
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...