Dried Blood and Serum Spots As A Useful Tool for Sample Storage to Evaluate Cancer Biomarkers
Published on: June 11, 2018
Measurement of MMP-7 in micro-volume peripheral blood: development of dried blood spot approach
Jingying Jiang1, Shuyang Liu2, Min Du1
1Department of Pediatric Surgery, Shanghai Key Laboratory of Birth Defect, and Key Laboratory of Neonatal Disease, Ministry of Health, Children's Hospital of Fudan University, Shanghai, China.
Purpose:
Serum matrix metalloproteinase-7 (MMP-7) is significant in differentiating biliary atresia (BA). This study aims to develop a new peripheral blood quantitative collection device to detect MMP-7 levels via dried blood spot (DBS).
Methods:
This is a diagnostic accuracy test. Serum and DBS MMP-7 concentrations were measured using an ELISA kit. Intraoperative cholangiography and subsequent histological examinations were used to confirm BA diagnoses.
Results:
A total of 241 infants with obstructive jaundice were enrolled, among whom 168 were BA. Linear regression showed DBS MMP-7 correlated well with serum MMP-7 (R = 0.93, P < 0.001). The best cut-off value of serum MMP-7 for BA was 25.9 ng/ml, achieving the area under the ROC curve (AUC) of 0.962 (95% CI: 0.941, 0.983), and the sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV) were 86.9%, 94.5%, 97.3% and 75.8%, respectively. The best cut-off value of DBS MMP-7 for BA was 12.5 ng/ml, achieving the AUC of 0.922 (95% CI: 0.888, 0.956), and the sensitivity, specificity, PPV, and NPV were 86.9%, 89.0%, 94.8%, and 74.7%, respectively. The dried blood spots were intervened under different storage conditions, including 1-5 days at room temperature, 2 or 3 days at 30 °C and 2 or 3 days at 37 °C. The DBS MMP-7 concentration under different storage conditions had good correlation and consistency with that at -80 °C.
Conclusions:
Serum and DBS MMP-7 correlate well, both of which have high accuracy in the diagnosis of BA, while the requirements for the storage of DBS are low.