Accuracy of iodine quantification in dual energy CT: A phantom study across 3 different CT systems
N Dehlinger1, J Bach1, T Willaume1
1Pole d'imagerie médicale, Hôpitaux universitaire de Strasbourg, Strasbourg, France.
Introduction:
No study has rigorously compared the performances of iodine quantification on recent CT systems employing different emission-based technologies, depending on the manufacturers and models.
Methods:
A specific bespoke phantom was used for this study, with 12 known concentrations of iodinated contrast agent: 0.4, 0.5, 1.0, 2.0, 3.0, 4.0, 5.0, 10.0, 15.0, 20.0, 30.0 and 50.0 mg/mL. Three different dual-energy scanners were tested: one system using dual-source acquisition (CT#1) and two systems using Fast kilovolt-peak switching technology ± artificial intelligence (AI) reconstruction methods (CT#2 and #3) from two different manufacturers. For each system, helical scans were performed following recommended clinical protocols. Four acquisitions were performed per iodine concentration (mg/mL), and measurements were made on iodine-maps using ROIs. Mean measured values were compared to the known concentrations, and the absolute quantification error (AQE) and the relative percentage error (RPE) were used to compare the performances of each CT.
Results:
The accuracy of the obtained measurements varied depending on the studied model but not on the acquisition mode (dual-source vs kVp switch ± AI). The quantification was more precise at high concentrations. RPE values were below 10 % with CT#2 (kVp switch) and below 25 % with CT#1 (dual-source), but were significantly higher with CT#3 (kVp switch + AI), exceeding 50 % at low concentrations (<3 mg/mL).
Conclusions:
With the help of a phantom, we identified variability in the results accuracy depending on the CT model, with sometimes significant deviation. Considering the performances of the different DECT technologies in iodine mapping, dual-source (CT#1) and kVp switch (CT#2) technologies appear more accurate than kVp switch technology combined with deep-learning-based reconstruction (CT#3) especially at low concentrations (<3 mg/mL).
Implications For Practice:
As the primary and daily user of medical imaging devices, the radiographer role is to be attentive to the performance of imaging systems, particularly when performing quantitative acquisitions like iodine-quantification. In CT quantitative imaging (iodine map), it's essential for radiographers to consider their CT systems as measuring tools, and to be aware of their accuracies and limits.
More Related Videos
08:41Novel In Vivo Micro-Computed Tomography Imaging Techniques for Assessing the Progression of Non-Alcoholic Fatty Liver Disease
Published on: March 24, 2023
08:59Use of a Multi-compartment Dynamic Single Enzyme Phantom for Studies of Hyperpolarized Magnetic Resonance Agents
Published on: April 15, 2016
Related Concept Videos
Radiological Investigation I: X-ray and CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET