Related Experiment Video
Updated: Jul 9, 2025

Hemodynamic Precision in the Neonatal Intensive Care Unit using Targeted Neonatal Echocardiography
Published on: January 27, 2023
Comparison of the effectiveness of different high-flow devices in neonatal care
Diogo Machado Kaminski1, Graziela Ferreira Biazus1, Rita C Silveira2
1Hospital de Clínicas de Porto Alegre, Department of Physiotherapy, Neonatology Service, Porto Alegre, RS, Brazil.
Objective:
To evaluate the success rate of high-flow nasal cannula (HFNC) therapy using an adapted obsolete mechanical ventilator (MV), Optiflow™ and Vapotherm™ in newborns (NBs).
Method:
This was a retrospective observational study conducted in the neonatal intensive care unit (NICU). The sample comprised NBs who underwent HFNC therapy due to ventilatory dysfunction, for weaning from non-invasive ventilation (NIV), or post-extubation. The three groups, stratified according to gestational age (GA) and birth weight, and corrected GA and weight at the beginning of HFNC use, were as follows: Optiflow ™, Vapotherm ™, and obsolete Mechanical Ventilator (MV) adapted for high flow therapy. Subsequently, the NBs were divided into a success group (SG) and a failure group (FG). HFNC success was defined as a therapy duration exceeding 72 h.
Results:
A total of 209 NBs were evaluated, with 31.1 % using HFNC due to ventilatory dysfunction, 2.4 % after extubation, and 66.5 % after NIV weaning. HFNC success rate was observed in 90.9 % of the NBs, with no difference between equipment types (Vapotherm ™, Optiflow ™, and adapted VM).
Conclusion:
Different types of HFNC equipment are equally effective when used in neonatology for respiratory dysfunction, as a method of weaning from NIV and post-extubation. Adapted obsolete MV can be an alternative for HFCN therapy in resource-constrained settings.
Related Concept Videos
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils,...