Comparisons of biometric parameters measurements by OPD-SCAN Ⅲ and Pentacam in cataract
Shiqi Dong1, Rong Xu1, Yue Wang1
1Department of Cataract, Hankou Aier Eye Hospital, Wuhan, Hubei Province, China.
Background:
Comparison of OPD-Scan III and Pentacam measurements of relevant parameters guiding the selection of multifocal intraocular lenses in cataract patients.
Methods:
A retrospective study included 120 patients (120 eyes) with cataract. Their chord kappa, chord alpha, corneal spherical aberration, and pupil size were measured by OPD-Scan Ⅲ and Pentacam. Pairwise t-tests, Pearson tests, and Bland-Altman analyses were used, respectively, to assess the difference, correlation, and agreement between the devices.
Results:
There was no significant difference in photopic chord kappa and corneal spherical aberration between the two instruments (P = 0.054, P = 0.065). Chord alpha and pupil size varied significantly between the two instruments (P < 0.001). OPD-Scan III results revealed a significant association between photopic chord kappa and mesopic chord kappa (r = 0.823, P < 0.001). There was a significant positive correlation between photopic and mesopic chord kappa measured by OPD-Scan Ⅲ and chord kappa measured by Pentacam (r = 0.840, r = 0.757, P < 0.001). The chord alpha evaluated by the two tools had a moderate correlation (r = 0.442, P < 0.001) between them. Bland-Altman analysis showed that there was excellent agreement between the two instruments in measuring chord kappa and corneal spherical aberration, while the chord alpha and pupil size were not consistent.
Conclusions:
The chord kappa and corneal spherical aberration measured by OPD-Scan III and Pentacam were consistent in the preoperative decision making of multifocal intraocular lens implantation in cataract patients, both of which have clinical guiding significance. In addition, OPD-Scan III provides more comprehensive day and night chord kappa and pupil diameters, which improves clinical advice for patients with complicated ocular diseases and high demands for nocturnal vision.