Related Experiment Video
Updated: Jul 9, 2025

Fabrication of Carbon-Based Ionic Electromechanically Active Soft Actuators
Published on: April 25, 2020
Feasibility of Parylene C for encapsulating piezoelectric actuators in active medical implants
Alina Kohler1,2,3, Felix Blendinger1, Sonja Müller1,4
1Faculty Mechanical and Medical Engineering (MME), Institute for Microsystems Technology (iMST), Furtwangen University, Furtwangen, Germany.
Abstract:
Parylene C is well-known as an encapsulation material for medical implants. Within the approach of miniaturization and automatization of a bone distractor, piezoelectric actuators were encapsulated with Parylene C. The stretchability of the polymer was investigated with respect to the encapsulation functionality of piezoelectric chips. We determined a linear yield strain of 1% of approximately 12-µm-thick Parylene C foil. Parylene C encapsulation withstands the mechanical stress of a minimum of 5×105 duty cycles by continuous actuation. The experiments demonstrate that elongation of the encapsulation on piezoelectric actuators and thus the elongation of Parylene C up to 0.8 mm are feasible.