Estimation of central blood pressure waveform from femoral blood pressure waveform by blind sources separation
Laila Gbaoui1, Christoph Hoeschen1, Eugenijus Kaniusas2
1Chair of Medical System Technology, Institute for Medical Instrumentation, Otto von Guericke University, Magdeburg, Germany.
Insights
A novel method accurately estimates central blood pressure (cBP) non-invasively from peripheral blood pressure (pBP) readings. This patient-specific approach identifies pressure waveform distortions, improving cardiovascular risk assessment.
Area of Science:
- Cardiovascular Physiology
- Biomedical Engineering
- Medical Diagnostics
Background:
- Central blood pressure (cBP) is a superior predictor of cardiovascular events compared to peripheral blood pressure (pBP).
- Current non-invasive cBP estimation relies on generalized transfer functions, lacking patient-specific accuracy.
- A validated, non-invasive method for personalized cBP derivation from pBP is crucial for clinical practice.
Purpose of the Study:
- To develop a novel blind source separation tool for non-invasive cBP estimation.
- To identify factors causing peripheral pressure waveform distortion.
- To create a patient-specific method without additional calibration or arterial model assumptions.
Main Methods:
- Employed single-channel independent component analysis (SCICA) to extract underlying pressure waveforms from femoral artery BP (BPfe).
- Utilized intra-aortic pressure (BPao) recordings from eight patients for validation.
- Evaluated SCICA accuracy using MAE, RMSE, RRMSE, ICC, regression, and Bland-Altman analyses for SBP, DBP, MAP, and PP.
Main Results:
- SCICA accurately estimated cBP waveforms from BPfe with high ICC (0.94) and low error metrics (MAE ±1.629 mmHg, RMSE ±0.957 mmHg).
- Strong linear correlations (R² > 0.97) were observed between estimated and intra-aortic SBP, DBP, MAP, and PP.
- Bland-Altman analysis confirmed good agreement between estimated and intra-aortic parameters, with minimal mean differences.
Conclusions:
- The SCICA approach effectively estimates central blood pressure non-invasively.
- This method successfully identifies sources of peripheral arterial pressure waveform distortion.
- SCICA offers a patient-specific, accurate tool for clinical cardiovascular risk assessment.
Background:
Central blood pressure (cBP) is a better indicator of cardiovascular morbidity and mortality than peripheral BP (pBP). However, direct cBP measurement requires invasive techniques and indirect cBP measurement is based on rigid and empirical transfer functions applied to pBP. Thus, development of a personalized and well-validated method for non-invasive derivation of cBP from pBP is necessary to facilitate the clinical routine. The purpose of the present study was to develop a novel blind source separation tool to separate a single recording of pBP into their pressure waveforms composing its dynamics, to identify the compounds that lead to pressure waveform distortion at the periphery, and to estimate the cBP. The approach is patient-specific and extracts the underlying blind pressure waveforms in pBP without additional brachial cuff calibration or any a priori assumption on the arterial model.
Methods:
The intra-arterial femoral BPfe and intra-aortic pressure BPao were anonymized digital recordings from previous routine cardiac catheterizations of eight patients at the German Heart Centre Berlin. The underlying pressure waveforms in BPfe were extracted by the single-channel independent component analysis (SCICA). The accuracy of the SCICA model to estimate the whole cBP waveform was evaluated by the mean absolute error (MAE), the root mean square error (RMSE), the relative RMSE (RRMSE), and the intraclass correlation coefficient (ICC). The agreement between the intra-aortic and estimated parameters including systolic (SBP), diastolic (DBP), mean arterial pressure (MAP), and pulse pressure (PP) was evaluated by the regression and Bland-Altman analyses.
Results:
The SCICA tool estimated the cBP waveform non-invasively from the intra-arterial BPfe with an MAE of 0.159 ± 1.629, an RMSE of 5.153 ± 0.957 mmHg, an RRMSE of 5.424 ± 1.304%, and an ICC of 0.94, as well as two waveforms contributing to morphological distortion at the femoral artery. The regression analysis showed a strong linear trend between the estimated and intra-aortic SBP, DBP, MAP, and PP with high coefficient of determination R2 of 0.98, 0.99, 0.99, and 0.97 respectively. The Bland-Altman plots demonstrated good agreement between estimated and intra-aortic parameters with a mean error and a standard deviation of difference of -0.54 ± 2.42 mmHg [95% confidence interval (CI): -5.28 to 4.20] for SBP, -1.97 ± 1.62 mmHg (95% CI: -5.14 to 1.20) for DBP, -1.49 ± 1.40 mmHg (95% CI: -4.25 to 1.26) for MAP, and 1.43 ± 2.79 mmHg (95% CI: -4.03 to 6.90) for PP.
Conclusions:
The SCICA approach is a powerful tool that identifies sources contributing to morphological distortion at peripheral arteries and estimates cBP.
More Related Videos
Related Concept Videos
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessing Blood pressure in the Leg
Preparation:
Assessment of blood pressure in brachial artery(two-step method)
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: