Background:
Ebstein's anomaly occurs when there is an apical displacement of the tricuspid valve with septal and posterior valve leaflets tethering. This condition often occurs in association with other congenital, structural, or conduction system diseases, including intracardiac shunts, valvular lesions, arrhythmias, accessory conduction pathways, and first-degree atrioventricular (AV) block. We present for the first time a case of a patient with Ebstein's anomaly who presented with second-degree Mobitz II AV block and was successfully treated with conduction system pacing (CSP) due to her young age and the likelihood of a long-term high percentage of pacing.
Case Summary:
We present a case of a 42-year-old lady with a background of complex congenital heart disease, including severe pulmonary stenosis, Ebstein anomaly, and atrial septal defect (ASD). She required complex surgical intervention, including tricuspid valve (TV) repair and subsequently replacement, ASD closure, and pulmonary balloon valvuloplasty. She presented to our hospital with symptomatic second-degree Mobitz II AV block (dizziness, shortness of breath, and exercise intolerance) and right bundle branch block (RBBB) on her baseline ECG. Her echocardiogram showed dilated right ventricle (RV) and left ventricle (LV) with low normal LV systolic function. Due to her young age and the likelihood of a long-term high percentage of RV pacing, we opted for CSP after a detailed discussion and patient consent. The distal HIS position is the preferred pacing strategy at our centre. We could not cross the TV with the standard Medtronic C315 HIS catheter, so we had to use the deflectable C304 HIS catheter. Mapping and pacing of the distal HIS bundle were achieved by Medtronic Selectsecure 3830, 69 cm lead. HIS bundle pacing led to the correction of both second-degree Mobitz II AV block and pre-existing RBBB. The implantation was uneventful, and the patient was discharged home the next day without any acute complications.
Discussion:
Distal HIS pacing is feasible in patients with surgically treated complex Ebstein anomaly and heart block. This approach can normalize the QRS complex with a high probability of preserving or improving LV function.
Related Concept Videos
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow...
Antiarrhythmic Drugs: Class IV Agents as Calcium Channel Blockers
Verapamil, a calcium channel blocker, inhibits calcium movement across myocardial cell membranes and vascular smooth muscle. This results in the dilation of coronary and...
Mechanism of Cardiac Arrhythmias
Antiarrhythmic Drugs: Class II Agents as β-Adrenergic Blockers