Risk factors for poor performance in finger cuff non-invasive monitoring of arterial pressure: A prospective
Karim Lakhal1, Jérôme E Dauvergne1, Hélène Messet-Charriere2
1Nantes Université, CHU Nantes, Department of Anesthesiology and Critical Care at Laënnec Hospital, University Hospital of Nantes, F-44000 Nantes, France.
Background:
Compared to the invasive technique, non-invasive monitoring of arterial pressure favors easier and faster implementation while potentially sacrificing some reliability. This may be particularly true for the Clearsight™ system (Edwards Lifesciences), which enables continuous monitoring. We evaluated the risk factors for its poor performance.
Methods:
Patients with an arterial catheter and stable mean arterial pressure (MAP) over a 5-min period were included. Six pairs of invasive and Clearsight measurements of MAP were collected and the bias between the two techniques was calculated. Poor performance of the Clearsight™ system was defined as either a failure to measure and display MAP or displaying an erroneous MAP (individual bias > 5 mmHg). Fingertip perfusion was assessed using the plethysmographic perfusion index (PI) and the capillary refill time (CRT).
Results:
Among 152 ICU patients (MAP of 81 ± 14 mmHg, norepinephrine in 78 [51%]), 78 (51%) experienced a poor performance of the Clearsight™ system: failure to display MAP in 19 (13%) patients, and erroneous value displayed in 59 (44%). In multivariate analysis, PI ≤ 0.85% (adjusted odds ratio [aOR] = 2.94 [95% confidence interval (95%CI):1.34;6.45]), CRT > 4 s (aOR = 5.28 [95%CI 1.39;20.05]), and the presence of hand edema (aOR = 2.06 [95%CI 1.01;4.21]) were associated with a higher likelihood of poor performance. Cardiac arrhythmia (aOR = 1.39 [95%CI 0.64;3.02]) and other tested variables were not associated with poor performance.
Conclusions:
Half of the included patients exhibited poor Clearsight™ system performance. Our results caution against using finger cuff arterial pressure monitoring in patients with low PI (≤0.85%), protracted CRT (>4 s), or hand edema.
Registration:
ClinicalTrials.gov, NCT04269382, Dr. G. Muller, February 13, 2020. https://classic.
Clinicaltrials:
gov/ct2/show/NCT04269382.
More Related Videos
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Pre-Procedural Guidelines for Assessing Blood Pressure
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure: