Precise detection and localization of R-peaks from ECG signals
Diguo Zhai1,2, Xinqi Bao3, Xi Long4
1Guangdong Provincial Key Laboratory of Optical Information Materials and Technology, Institute of Electronic Paper Displays, South China Academy of Advanced Optoelectronics, South China Normal University, Guangzhou 510006, China.
Abstract:
Heart rate variability (HRV) is derived from the R-R interval, which depends on the precise localization of R-peaks within an electrocardiogram (ECG) signal. However, current algorithm assessment methods prioritize the R-peak detection's sensitivity rather than the precision of pinpointing the exact R-peak positions. As a result, it is of great value to develop an R-peak detection algorithm with high-precision R-peak localization. This paper introduces a novel R-peak localization algorithm that involves modifications to the well-established Pan-Tompkins (PT) algorithm. The algorithm was implemented as follows. First, the raw ECG signal $ X\left(i\right) $ was band-pass filtered (5-35 Hz) to obtain a preprocessed signal $ Y\left(i\right) $. Second, $ Y\left(i\right) $ was squared to enhance the QRS complex, followed by a 5 Hz low-pass filter to obtain the QRS envelope, which was transformed into a window signal $ W\left(i\right) $ by dynamic threshold with a minimum width of 200 ms to mark the QRS complex. Third, $ Y\left(i\right) $ was used to generate QRS template $ T\left(n\right) $ automatically, and then the R-peak was identified by a template matching process to find the maximum absolute value of all cross-correlation values between $ T\left(n\right) $ and $ Y\left(i\right) $. The proposed algorithm achieved a sensitivity (SE) of 99.78%, a positive prediction value (PPV) of 99.78% and data error rate (DER) of 0.44% in R-peak localization for the MIT-BIH Arrhythmia database. The annotated-detected error (ADE), which represents the error between the annotated R-peak location and the detected R-peak location, was 8.35 ms for the MIT-BIH Arrhythmia database. These results outperformed the results obtained using the classical Pan-Tompkins algorithm which yielded an SE of 98.87%, a PPV of 99.14%, a DER of 1.98% and an ADE of 21.65 ms for the MIT-BIH Arrhythmia database. It can be concluded that the algorithm can precisely detect the location of R-peaks and may have the potential to enhance clinical applications of HRV analysis.
Related Concept Videos
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...
ECG Interpretation of Rhythms
Components of the Electrocardiogram
The primary components of a normal ECG waveform in Normal sinus rhythm(NSR) include the P wave, PR interval, QRS complex, ST segment, T wave, and occasionally a U wave.
ECG waveforms are divided by vertical and horizontal lines at standard intervals.
The horizontal axis measures time and rate, and the vertical axis measures amplitude or voltage....
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin...
Instrumentation Amplifier
To overcome this challenge, an ECG machine utilizes an instrumentation amplifier. This specialized amplifier is...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Correlation between ECG and Cardiac Cycle
A cardiac action potential originates in the SA node and spreads throughout the atria and the AV node in approximately 0.03 seconds. This results in the P wave in an ECG and triggers atrial contraction. The action potential is then briefly slowed at the AV node, allowing the atria to contract and fill the ventricles with blood before...