Ensuring quality in 17OHP mass spectrometry measurement: an international study assessing isomeric steroid
Chung Shun Ho1, Kirsten Hoad2, Brian R Cooke2
1Department of Chemical Pathology, Biomedical Mass Spectrometry Unit, The Chinese University of Hong Kong, Prince of Wales Hospital, Shatin, New Territories, Hong Kong SAR China.
Objectives:
Interference from isomeric steroids is a potential cause of disparity between mass spectrometry-based 17-hydroxyprogesterone (17OHP) results. We aimed to assess the proficiency of mass spectrometry laboratories to report 17OHP in the presence of known isomeric steroids.
Methods:
A series of five samples were prepared using a previously demonstrated commutable approach. These samples included a control (spiked to 15.0 nmol/L 17OHP) and four challenge samples further enriched with equimolar concentrations of 17OHP isomers (11α-hydroxyprogesterone, 11β-hydroxyprogesterone, 16α-hydroxyprogesterone or 21-hydroxyprogesterone). These samples were distributed to 38 participating laboratories that reported serum 17OHP results using mass spectrometry in two external quality assurance programs. The result for each challenge sample was compared to the control sample submitted by each participant.
Results:
Twenty-six laboratories (68 % of distribution) across three continents returned results. Twenty-five laboratories used liquid chromatography-tandem mass spectrometry (LC-MS/MS), and one used gas chromatography-tandem mass spectrometry to measure 17OHP. The all-method median of the control sample was 14.3 nmol/L, ranging from 12.4 to 17.6 nmol/L. One laboratory had results that approached the lower limit of tolerance (minus 17.7 % of the control sample), suggesting the isomeric steroid caused an irregular result.
Conclusions:
Most participating laboratories demonstrated their ability to reliably measure 17OHP in the presence of the four clinically relevant isomeric steroids. The performance of the 12 (32 %) laboratories that did not engage in this activity remains unclear. We recommend that all laboratories offering LC-MS/MS analysis of 17OHP in serum, plasma, or dried bloodspots determine that the isomeric steroids are appropriately separated.
More Related Videos
10:17High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
08:53Quantitative and Qualitative Method for Sphingomyelin by LC-MS Using Two Stable Isotopically Labeled Sphingomyelin Species
Published on: May 7, 2018
Related Concept Videos
High-Resolution Mass Spectrometry (HRMS)
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Mass Spectrometry: Isotope Effect
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
Mass Spectrum: Interpretation
To...