Development of in-vitro pulsatile flow generator for evaluating the performance of hemodialysis catheters
Seongsu Cho1, Jihyeong Lee1, Sun Cheol Park2
1School of Mechanical Engineering, Sungkyunkwan University, Suwon 16419, Republic of Korea.
Abstract:
Hemodialysis (HD) using an HD catheter is performed widely on renal failure patients. The catheter was evaluated using the recirculation ratio in pre-clinical status, which is a crucial index indicating its performance. However, pre-clinical in-vivo experiments have limitations: high cost, and ethical issues. Hence, computational and in-vitro methods have been developed as alternatives. However, computational methods require fluid dynamic knowledge, whereas in-vitro experiments are complicated and expensive. In this study, we developed a pulsatile flow generator to mimic blood flow achieving cost effectiveness and user convenience. The device used iterative learning control, achieving blood flow in the superior and inferior vena cava within a 3.3% error. Furthermore, the recirculation ratios were measured based on two insertion directions and two different external pipe materials to evaluate the catheter regarding patients' posture and blood vessel stiffness. The results provide a better understanding of cardiovascular device performance without complicated and costly pre-clinical tests.
More Related Videos
07:30In Vitro Model of Physiological and Pathological Blood Flow with Application to Investigations of Vascular Cell Remodeling
Published on: November 3, 2015
08:44An In Vitro Hemodynamic Loop Model to Investigate the Hemocytocompatibility and Host Cell Activation of Vascular Medical Devices
Published on: August 21, 2020
Related Concept Videos
Sleep Apnea
The condition is more prevalent among...
Cardiopulmonary Resuscitation II: ACLS Airway Management