A diagnostic to measure time-resolved atom column density and Doppler temperature in atomic vapors produced by laser
M M Schauer1, J A Paisner1, G D Stevens2
1Los Alamos National Laboratory, Los Alamos, New Mexico 87545, USA.
Abstract:
We report on the development of a diagnostic to measure the time-resolved column density and Doppler temperature of atomic vapors produced by laser ablation. The diagnostic is based on the strong frequency dependence of the atomic susceptibility near an electronic transition in the interrogated atomic species. Interference on the face of a fast photodetector between the several frequency components present in a sinusoidally phase-modulated probe beam will produce a time signature uniquely determined by the column density of atoms in the probed atomic state and the Doppler temperature of the atomic vapor. With the extensive, high precision atomic spectroscopy data available in the literature, it is possible to model the vapor and extract the desired parameters through comparison of the model result with the experimental data.
Related Concept Videos
Atomic Absorption Spectroscopy: Atomization Methods
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing...
Atomic Spectroscopy: Effects of Temperature
At thermal equilibrium, the relative populations of excited and ground state atoms can be estimated using the Maxwell–Boltzmann distribution. For example, an increase in temperature...
Atomic Emission Spectroscopy: Lab
Atomic Absorption Spectroscopy: Instrumentation
The atomizer used in AAS can be either a flame atomizer or an...
Atomic Spectroscopy: Absorption, Emission, and Fluorescence