Related Experiment Video
Updated: Jul 8, 2025

Fabrication of High Contact-Density, Flat-Interface Nerve Electrodes for Recording and Stimulation Applications
Published on: October 4, 2016
A Multielectrode Nerve Cuff for Chronic Velocity Selective Recording in a sheep model
Abstract:
Supra-sacral spinal cord injury (SCI) causes loss of bladder fullness sensation and bladder over-activity, leading to retention and incontinence respectively. Velocity selective recording (VSR) of nerve roots innervating the bladder might enable identification of bladder activity. A 10-electrode nerve cuff for sacral nerve root VSR was developed and tested in a sheep model during acute surgeries and chronic implantation for 6 months. The cuff performed well, with 5.90±1.90 kΩ electrode, and <~800 Ω tissue impedance after 189 days implantation with a stable device and tissues. This is important information for assessing the feasibility of chronic VSR.Clinical Relevance-This demonstrates the manufacturing and performance of a neural interface for chronic monitoring of bladder nerve afferents with applications in urinary incontinence and retention management following SCI.