Summary
This study developed a nerve cuff for velocity selective recording (VSR) to monitor bladder activity after spinal cord injury (SCI). The device showed stable performance for chronic VSR, aiding bladder control management.
Area of Science:
- Neuroscience
- Biomedical Engineering
- Urology
Background:
- Supra-sacral spinal cord injury (SCI) disrupts bladder control, causing sensory loss and overactivity.
- Current management for SCI-related bladder dysfunction is limited.
- Velocity selective recording (VSR) offers potential for monitoring bladder nerve activity.
Purpose of the Study:
- To develop and evaluate a 10-electrode nerve cuff for chronic VSR of sacral nerve roots.
- To assess the feasibility of using VSR for monitoring bladder afferents in a chronic model.
- To demonstrate a neural interface for managing urinary incontinence and retention post-SCI.
Main Methods:
- A 10-electrode nerve cuff was designed for sacral nerve root VSR.
- The device was tested in a sheep model during acute surgery and chronic implantation (6 months).
- Electrode impedance and tissue impedance were measured to assess device performance and stability.
Main Results:
- The nerve cuff demonstrated good performance with stable electrode impedance (5.90±1.90 kΩ).
- Low tissue impedance (<~800 Ω) was maintained after 189 days of implantation.
- The device and surrounding tissues remained stable throughout the chronic implantation period.
Conclusions:
- The developed nerve cuff is suitable for chronic VSR of sacral nerve roots.
- This technology shows promise for long-term monitoring of bladder nerve activity.
- The neural interface is a viable tool for managing urinary incontinence and retention after SCI.