Simultaneous determination of cystine and other free aminothiols in blood plasma using capillary electrophoresis with pH-mediated stacking
Alexander Vladimirovich Ivanov1, Mikhail Aleksandrovich Popov2, ValeryVasil'evich Aleksandrin1
1Department of Molecular and Cell Pathophysiology, Federal State Budgetary Scientific Institution 'Institute of General Pathology and Pathophysiology', Moscow, Russia.
Abstract:
We developed a method of sensitive capillary electrophoresis using UV detection for the determination of certain free aminothiols (reduced cysteinylglycine (rCysGly), cysteine (rCys), glutathione (rGln), and cystine (CysS) in human blood plasma. The reduced thiols were derivatized with N-ethylmaleimide. The plasma was purified from proteins via ultrafiltration. Electrophoretic separation was performed using 115 mM Na phosphate with 7.5% (v/v) polyethylene glycol 600, pH 2.3. The in-capillary concentration of the analytes was achieved with a pH gradient created via the preinjection of triethanolamine and postinjection of phosphoric acid. The separation was carried out using a silica capillary (50 µm i.d.; total/effective separation length 42/35 cm) at a 25 kV voltage. The total analysis/regeneration time was 18 min. The quantification limits varied from 1.3 µM (rCysGly) to 5.4 µM (CysS). The accuracy was 95%-99%, and the repeatability and reproducibility were approximately 1.8%-3.8% and 1.9%-5.0%, respectively. An analysis of plasma samples from healthy volunteers (N = 41) showed that the mean levels of rCysGly, rCys, rGln, and CysS were 1.64, 10.6, 2.58, and 46.2 µM, respectively.
More Related Videos
Related Concept Videos
Capillary Electrophoresis: Instrumentation
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...
Electrophoresis: Overview
There...
SDS-PAGE
A variation of gel electrophoresis, termed polyacrylamide gel electrophoresis (PAGE), is commonly used for separating proteins according to their molecular size by passing them through a polyacrylamide gel. Because of the varying charges associated with amino acid side chains, PAGE can be used to separate intact...