Purpose Of Review:
Accumulating evidence ascribes the benefit of extracorporeal gas exchange, at least in most severe cases, to the provision of a lung healing environment through the mitigation of ventilator-induced lung injury (VILI) risk. In spite of pretty homogeneous criteria for extracorporeal gas exchange application (according to the degree of hypoxemia/hypercapnia), ventilatory management during extracorporeal membrane oxygenation (ECMO)/carbon dioxide removal (ECCO 2 R) varies across centers. Here we summarize the recent evidence regarding the management of mechanical ventilation during extracorporeal gas exchange for respiratory support.
Recent Findings:
At present, the most common approach to protect the native lung against VILI following ECMO initiation involves lowering tidal volume and driving pressure, making modest reductions in respiratory rate, while typically maintaining positive end-expiratory pressure levels unchanged.Regarding ECCO 2 R treatment, higher efficiency devices are required in order to reduce significantly respiratory rate and/or tidal volume.
Summary:
The best compromise between reduction of native lung ventilatory load, extracorporeal gas exchange efficiency, and strategies to preserve lung aeration deserves further investigation.
Related Concept Videos
Acute Respiratory Failure-V
Ensure that patients are monitored continuously for their response to therapy, including changes in...
Acute Respiratory Failure-III
Acute Respiratory Failure-I
Definition: It is defined by specific criteria based on blood gas measurements. Hypoxemia happens when the partial pressure of oxygen (PaO2) falls below 60 mmHg. At the same time,...
Acute Respiratory Failure-II
The underlying physiological abnormalities that contribute to hypoxemic respiratory failure include: