Related Experiment Video
Updated: Jul 8, 2025

Author Spotlight: Advancing Endoscopic Ossiculoplasty – Techniques, Innovations, and Practical Guidance for Clinical Integration
Published on: January 26, 2024
[Discussion of Application Status and Key Attentions on Registration for Ossicular Replacement Prosthesis]
Jing Wu1, Jiazhen Zhang1, Ruhan A1
1Center for Medical Device Evaluation, NMPA, Beijing, 100081.
Abstract:
The ossicular replacement prosthesis should have good biocompatibility, stability, easy to install, and excellent sound transmission capacity. In this study, the characteristics of ideal materials for the ossicular replacement prosthesis were analyzed by searching the types of materials used in clinical practice and comparing the advantages and disadvantages of various materials and structures. At the same time, in combination with the current evaluation requirements and evaluation experience, the focus of the performance research project of ossicular replacement prosthesis in the process of registration is discussed to clarify the performance evaluation requirements of these products, so as to provide reference for the future work of manufacturers and regulators. The performance evaluation of ossicular replacement prosthesis focuses on its mechanical properties, fixation stability, sound transmission characteristics, biological characteristics, and magnetic resonance compatibility.