Related Experiment Video
Updated: Jul 8, 2025

Precision Measurements and Parametric Models of Vertebral Endplates
Published on: September 17, 2019
Applying Machine Learning and Point-Set Registration to Automatically Measure the Severity of Spinal Curvature on
Jason Wong1, Marek Reformat1, Edmond Lou1
1Department of Electrical and Computer EngineeringUniversity of Alberta Edmonton AB T6G 1H9 Canada.
Objective:
Measuring the severity of the lateral spinal curvature, or Cobb angle, is critical for monitoring and making treatment decisions for children with adolescent idiopathic scoliosis (AIS). However, manual measurement is time-consuming and subject to human error. Therefore, clinicians seek an automated measurement method to streamline workflow and improve accuracy. This paper reports on a novel machine learning algorithm of cascaded convolutional neural networks (CNN) to measure the Cobb angle on spinal radiographs automatically.
Methods:
The developed method consisted of spinal column segmentation using a CNN, vertebra localization and segmentation using iterative vertebra body location coupled with another CNN, point-set registration to correct vertebra segmentations, and Cobb angle measurement using the final segmentations. Measurement performance was evaluated with the circular mean absolute error (CMAE) and percentage within clinical acceptance ([Formula: see text]) between automatic and manual measurements. Analysis was separated by curve severity to identify any potential systematic biases using independent samples Student's t-tests.
Results:
The method detected 346 of the 352 manually measured Cobb angles (98%), with a CMAE of 2.8° and 91% of measurements within the 5° clinical acceptance. No statistically significant differences were found between the CMAEs of mild ([Formula: see text]), moderate (25°-45°), and severe ([Formula: see text]) groups. The average measurement time per radiograph was 17.7±10.2s, improving upon the estimated average of 30s it takes an experienced rater to measure. Additionally, the algorithm outputs segmentations with the measurement, allowing clinicians to interpret measurement results.
Discussion/Conclusion:
The developed method measured Cobb angles on radiographs automatically with high accuracy, quick measurement time, and interpretability, suggesting clinical feasibility.