Expired CO2 Measurement in Intubated or Spontaneously Breathing Patients from the Emergency Department
Published on: January 29, 2011
Comparing the EMMA capnograph with sidestream capnography and arterial carbon dioxide pressure at 284 kPa
Xavier Ce Vrijdag1,2, Hanna van Waart1, Chris Sames3
1Department of Anaesthesiology, University of Auckland, Auckland, New Zealand.
Introduction:
Capnography aids assessment of the adequacy of mechanical patient ventilation. Physical and physiological changes in hyperbaric environments create ventilation challenges which make end-tidal carbon dioxide (ETCO2) measurement particularly important. However, obtaining accurate capnography in hyperbaric environments is widely considered difficult. This study investigated the EMMA capnograph for hyperbaric use.
Methods:
We compared the EMMA capnograph to sidestream capnography and the gold standard arterial carbon dioxide blood gas analysis in a hyperbaric chamber. In 12 resting subjects breathing air at 284 kPa, we recorded ETCO2 readings simultaneously derived from the EMMA and sidestream capnographs during two series of five breaths (total 24 measurements). An arterial blood gas sample was also taken simultaneously in five participants.
Results:
Across all measurements there was a difference of about 0.1 kPa between the EMMA and sidestream capnographs indicating a very slight over-estimation of ETCO2 by the EMMA capnograph, but fundamentally good agreement between the two end-tidal measurement methods. Compared to arterial blood gas pressure the non-significant difference was about 0.3 and 0.4 kPa for the EMMA and sidestream capnographs respectively.
Conclusions:
In this study, the EMMA capnograph performed equally to the sidestream capnograph when compared directly, and both capnography measures gave clinically acceptable estimates of arterial PCO2.
More Related Videos
07:51Evaluation of Capnography Sampling Line Compatibility and Accuracy when Used with a Portable Capnography Monitor
Published on: September 29, 2020
08:25Combining Volumetric Capnography And Barometric Plethysmography To Measure The Lung Structure-function Relationship
Published on: January 8, 2019
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Diagnosing Acidosis and Alkalosis
First, the pH level is assessed to determine whether the blood pH is normal (7.35–7.45), low (acidosis), or high (alkalosis).
Next, the PCO2 and...
Blood Studies I: ABG and VBG
Arterial Blood Gas (ABG)
Arterial Blood Gas (ABG) studies are crucial for assessing the lungs' ability to supply oxygen and remove carbon dioxide, reflecting the patient's ventilation status. They also help understand the kidneys' capacity to...
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Acute Respiratory Failure-III
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...