Related Experiment Video
Updated: Jul 27, 2026

Evaluation of Biomaterials for Bladder Augmentation using Cystometric Analyses in Various Rodent Models
Published on: August 9, 2012
Agreement and Reliability of Patient-measured Postvoid Residual Bladder Volumes
Michael Jalfon1, Mursal Gardezi1, Dylan Heckscher1
1Yale University School of Medicine, New Haven, CT.
Patients can reliably self-measure postvoid residual bladder volume (PVR) using portable ultrasound devices. This method is feasible, accurate, and preferred by patients for remote monitoring of voiding dysfunction.
Area of Science:
- Urology
- Medical Devices
- Remote Patient Monitoring
Background:
- Postvoid residual bladder volume (PVR) measurement is crucial for managing voiding dysfunction.
- Current PVR measurement often requires healthcare visits, limiting remote monitoring capabilities.
- Smart-device portable ultrasound offers a potential solution for at-home PVR assessment.
Purpose of the Study:
- To evaluate the reliability and agreement of patient self-measurement of PVR using a portable ultrasound system.
- To compare patient self-measurement with provider measurements.
- To assess patient preferences for self-measurement versus provider measurement.
Main Methods:
- Fifty patients undergoing urology outpatient visits participated.
- Providers and patients used a portable ultrasound device (in two modes) and a standard bladder scanner for PVR measurements.
- Reliability was assessed using intraclass correlation, and agreement using Bland-Altman analyses.
Main Results:
- Patient self-measurement demonstrated high reliability (intraclass correlations 0.95–0.98).
- Limits of agreement between self-measurement and standard scanner exceeded the 50 mL threshold.
- A majority of patients (74%) preferred self-measurement over provider measurement.
Conclusions:
- Patient self-measurement of PVR with portable ultrasound is feasible, reliable, and preferred.
- While agreement limits require consideration, this technology supports remote monitoring for voiding dysfunction.
- Further research should account for inherent ultrasound measurement errors in longitudinal studies.
Related Concept Videos
Respiratory Volumes
Tidal Volume (TV) Tidal volume (TV) is the air inhaled or exhaled in a...
Nursing Assessment of the Genitourinary System II: Inspection and Palpation
Imaging Studies II: Ultrasonography
Imaging Studies V: Intravenous Urography and Retrograde Pyelography
Imaging Studies VI: Voiding Cystourethrography and Cystography
Urodynamic Studies: Uroflowmetry