Force degradation of two orthodontic accessories analyzed in vivo and in vitro
Liu Yang1, Chenxing Lv2, Xiaomin Li3
1Department of Stomatology, Hangzhou Traditional Chinese Medicine Hospital Affiliated to Zhejiang Chinese Medical University, Hangzhou, China.
Objective:
To compare force degradation of elastomeric chains and NiTi coil springs in vivo and in vitro, and evaluate the effects of pre-stretched and reused elastomeric chains in the oral cavity during the time.
Methods:
In the in vitro groups, 4-unit elastomeric chains and NiTi coil springs with an initial force of 200 g were placed in dry air and artificial saliva. The volunteers wore clear retainers which were used to hold the sample of 4-unit chains, pre-stretched 4-unit chains, and NiTi coil springs with the initial force of 200 g in the in vivo groups. After the first 4 weeks, 4-unit specimens were stretched to 200 g again for another 4 weeks in vivo. The force value and the percentage of force degradation were recorded at each measurement time interval in the in vivo and in vitro groups.
Results:
The force degradation of elastomeric chains was greatest within the initial 4 hours, followed by a more stable phase after 1 week. The average force degradation of 4-unit elastomeric chains after 4 weeks was in vivo (64.8%) > artificial saliva (55.0%) > dry air (46.42%) (P < 0.05). The force degradation of NiTi coil springs in vivo (15.36%) or in artificial saliva (15.8%) was greater than in dry air (7.6%) (P < 0.05). NiTi coil springs presented a gentler force decay than elastomeric chains during the period (P < 0.05). In vivo, the force degradation of pre-stretched and reused elastomeric chains decreased less than the regular style(P < 0.05).
Conclusion:
The force degradation of the elastomeric chains and NiTi coil springs varied in different environments. NiTi coil springs presented a gentler force decay than elastomeric chains during the period. Orthodontists should consider the force degradation characteristics of orthodontic accessories in clinical practice.