Insights
Handheld ultrasound devices significantly reduced screen-positive rates for rheumatic heart disease (RHD) in Ethiopian children, indicating device choice impacts screening effectiveness.
Area of Science:
- Cardiology
- Public Health
- Medical Imaging
Background:
- Rheumatic heart disease (RHD) impacts 33 million globally, causing significant cardiovascular mortality in young individuals.
- Penicillin prophylaxis is key for managing asymptomatic RHD progression.
- Portable and handheld ultrasound devices are increasingly vital for expanding RHD screening, especially with simplified protocols.
Purpose of the Study:
- To evaluate the impact of handheld ultrasound devices on RHD screening outcomes in Ethiopia.
- To compare the performance of handheld ultrasound devices against traditional devices in RHD screening.
Main Methods:
- A retrospective analysis compared screen-positive ultrasound rates before and after introducing handheld ultrasound in an Ethiopian RHD screening program.
- A cross-sectional device comparison was conducted on 19 at-risk schoolchildren.
Main Results:
- Screen-positive rates decreased from 11.2% to 4.1% after handheld device implementation (p<0.001).
- Significant differences in interpretation were noted between expert and non-expert reviews with the Lumify device (p=0.025).
- A trend towards shorter jet length by color Doppler was observed with handheld devices.
Conclusions:
- The choice of ultrasound device significantly influences the screen-positive rate in RHD screening programs.
- Handheld ultrasound technology shows promise for improving RHD screening efficiency and potentially reducing false positives.
Abstract:
Background: Rheumatic heart disease (RHD) affects 33 million people in low and middle income countries and is the leading cause of cardiovascular death among children and young adults. Penicillin prophylaxis prevents progression in asymptomatic disease. Efforts to expand echocardiographic screening are focusing on simplified protocols, non-physician ultrasonographers, and portable ultrasound devices, including handheld ultrasound. Recent advances support the use of single-view screening protocols. With the increasing availability and low cost of handheld devices, studies are needed to evaluate their performance in these settings. Methods: We conducted a retrospective study comparing the rate of screen positive ultrasounds before and after the use of a handheld ultrasound in an RHD screening program in Ethiopia. We also performed a cross-sectional device comparison in 19 at-risk school-children participating in the rheumatic heart disease screening program. Results: Between March of 2019 and January of 2022, 6631 children were screened for rheumatic heart disease of whom 4029 were screened after the introduction of a handheld device. Before the use of the handheld ultrasound device 291 (11.2%) children had a screen positive ultrasounds compared with 167 (4.1%) afterwards (p<0.001). We also compared non-expert to expert interpretation by device and found a significant difference in interpretation for the Lumify (p=0.025). There was a trend towards shorter jet length by color Doppler in the handheld ultrasound device for both expert and non-expert review. Conclusions: Our study highlights that the screen-positive rate in a RHD screening program is influenced by the device being used in the screening process.
Related Concept Videos
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Ultrasonography
During an ultrasonography procedure, a handheld device called...
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion,...