The Impact of a Handheld Ultrasound Device in a Rheumatic Heart Disease Screening Program in Ethiopia

Zachary P Kaltenborn1, Anteneh Zewde2, Jonathan D Kirsch2

  • 1Departments of General Internal Medicine and Pediatrics, University of Minnesota Medical School Minneapolis , MN USA.

POCUS Journal
|December 15, 2023
PubMed

Insights

Handheld ultrasound devices significantly reduced screen-positive rates for rheumatic heart disease (RHD) in Ethiopian children, indicating device choice impacts screening effectiveness.

Area of Science:

  • Cardiology
  • Public Health
  • Medical Imaging

Background:

  • Rheumatic heart disease (RHD) impacts 33 million globally, causing significant cardiovascular mortality in young individuals.
  • Penicillin prophylaxis is key for managing asymptomatic RHD progression.
  • Portable and handheld ultrasound devices are increasingly vital for expanding RHD screening, especially with simplified protocols.

Purpose of the Study:

  • To evaluate the impact of handheld ultrasound devices on RHD screening outcomes in Ethiopia.
  • To compare the performance of handheld ultrasound devices against traditional devices in RHD screening.

Main Methods:

  • A retrospective analysis compared screen-positive ultrasound rates before and after introducing handheld ultrasound in an Ethiopian RHD screening program.
  • A cross-sectional device comparison was conducted on 19 at-risk schoolchildren.

Main Results:

  • Screen-positive rates decreased from 11.2% to 4.1% after handheld device implementation (p<0.001).
  • Significant differences in interpretation were noted between expert and non-expert reviews with the Lumify device (p=0.025).
  • A trend towards shorter jet length by color Doppler was observed with handheld devices.

Conclusions:

  • The choice of ultrasound device significantly influences the screen-positive rate in RHD screening programs.
  • Handheld ultrasound technology shows promise for improving RHD screening efficiency and potentially reducing false positives.

Related Concept Videos

Assessing Blood pressure using a doppler ultrasound01:19

Assessing Blood pressure using a doppler ultrasound

To obtain accurate blood pressure measurements in clinical settings, especially when traditional methods are insufficient, healthcare professionals utilize the Doppler ultrasound technique. This method uses high-frequency sound waves to detect blood flow within the arteries, which is crucial for patients with conditions that complicate circulatory system assessment.
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
1.4K
Ultrasonography01:17

Ultrasonography

Ultrasonography is an imaging technique that uses high-frequency sound waves to visualize the body's internal structures. It is a non-invasive and safe procedure that does not involve the use of ionizing radiation, making it widely used in various medical fields. Ultrasonography is used to study heart function, blood flow in the neck or extremities, certain conditions such as gallbladder disease, and fetal growth and development.
During an ultrasonography procedure, a handheld device called...
4.5K
Ultrasound II: Endoscopic Ultrasound and FibroScan01:25

Ultrasound II: Endoscopic Ultrasound and FibroScan

Endoscopic Ultrasound (EUS) and FibroScan are valuable diagnostic tools in gastroenterology and hepatology, each with specific applications and techniques.
Endoscopic Ultrasound (EUS):
111
Imaging Studies for Cardiovascular System I:Echocardiography01:17

Imaging Studies for Cardiovascular System I:Echocardiography

Cardiac imaging studies encompass a wide range of noninvasive and minimally invasive techniques designed to visualize the heart's structure and function in detail. One such technique is echocardiography, which uses high-frequency ultrasound waves to produce detailed images of the heart, known as echocardiograms.
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion,...