Related Experiment Video
Updated: Jul 21, 2026

An Intramedullary Locking Nail for Standardized Fixation of Femur Osteotomies to Analyze Normal and Defective Bone Healing in Mice
Published on: November 13, 2016
Biomechanical comparison of different fixation methods in tibiotalocalcaneal arthrodesis: a cadaver study
Matthias Trost1, Suzan Yarkin2, Matthias Knieps3
1Department of Orthopaedics and Traumatology, Saint Josef Hospital, Ruhr University Bochum, Gudrunstrasse 56, 44791, Bochum, Germany. trost@t-online.de.
Background:
Various fixation methods are available for tibiotalocalcaneal arthrodesis: nail, plate, or screws. An intramedullary bone stabilization system within a balloon catheter has not previously been used in tibiotalocalcaneal arthrodesis. The aim of this study was to compare the stability of these techniques.
Methods:
Twenty-four lower legs from fresh-frozen human cadavers were used. Tibiotalocalcaneal arthrodesis was performed with a retrograde nail, a lateral locking plate, three cancellous screws, or an intramedullary bone stabilization system. The ankles were loaded cyclically in plantarflexion and dorsiflexion.
Results:
For cyclic loading at 125 N, the mean range of motion was 1.7 mm for nail, 2.2 mm for plate, 6.0 mm for screws, and 9.0 mm for the bone stabilization system (P < .01). For cyclic loading at 250 N, the mean range of motion was 4.4 mm for nail, 7.5 mm for plate, 12.1 mm for screws, and 14.6 mm for the bone stabilization system (P < .01). The mean cycle of failure was 4191 for nail, 3553 for plate, 3725 for screws, and 2132 for the bone stabilization system (P = .10).
Conclusions:
The stability of the tibiotalocalcaneal arthrodesis differs depending on the fixation method, with nail or plate showing the greatest stability and the bone stabilization system the least. When three screws are used for tibiotalocalcaneal arthrodesis, the stability is intermediate. As the biomechanical stability of the bone stabilization system is low, it cannot be recommended for tibiotalocalcaneal arthrodesis.