Performance of wearable watch-type home blood pressure measurement devices in a real-world clinical sample
Mattia Lunardi1,2,3, Farooq Muhammad2, Atif Shahzad2
1Department of Cardiology, Saolta Group, Galway University Hospital, Health Service Executive and University of Galway, Galway, H91 TK33, Ireland.
Background:
Independent testing of home blood pressure (BP) measurement (HBPM) devices is often lacking, particularly among older and multi-morbid patients.
Methods:
We studied the Bpro G2 (using tonometry), Omron HeartGuide (using occlusive oscillometric technology), and Heartisans (using photoplethysmography) wrist watch HBPM devices against a gold standard brachial sphygmomanometer. To test device performance, we used the ISO81060-2 protocol (though this protocol cannot formally validate cuffless devices). We also used linear mixed models to compare adjusted longitudinal BP measurements between devices. Finally, as a surrogate for usability, we recorded instances of device failure where no BP measurement was returned.
Results:
We enrolled 128 participants (median [Q1-Q3] age 53 [40-65] years, 51% male, 46% on antihypertensive drugs), of whom 100 were suitable for the primary analysis. All three devices had mean BP values within 5 mmHg of sphygmomanometry. However, due to insufficient reliability (e.g., wider than accepted standard deviations of mean BP), none of the three devices passed all criteria required by the ISO81060-2 protocol. In adjusted longitudinal analyses, the Omron device also systematically underestimated systolic and diastolic BP (- 8.46 mmHg; 95% CI 6.07, 10.86; p < 0.001; and - 2.53 mmHg; 95% CI - 4.03, - 1.03; p = 0.001; respectively). Nevertheless, compared to the Omron device, BPro and Heartisans devices had increased odds of failure (BPro: odds ratio [OR] 5.24; p < 0.0001; Heartisans: OR 5.61; p < 0.001).
Conclusions:
While we could not formally validate the cuffless devices, our results show that wearable technologies will require improvements to offer reliable BP assessment. This study also highlights the need for validation protocols specifically designed for cuffless BP measurement technologies.
More Related Videos
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Errors occurring during blood pressure monitoring
Several factors...
Measurement of Blood Pressure
Pre-Procedural Guidelines for Assessing Blood Pressure