Understanding Error Patterns: An Analysis of Alignment Errors in Rigid 3D Body Scans
Julian Meißner1,2, Michael Kisiel1, Nagarajan M Thoppey1
1BSN Medical GmbH, Schützenstraße 1-3, 22761 Hamburg, Germany.
Abstract:
Three-dimensional body scanners are attracting increasing interest in various application areas. To evaluate their accuracy, their 3D point clouds must be compared to a reference system by using a reference object. Since different scanning systems use different coordinate systems, an alignment is required for their evaluation. However, this process can result in translational and rotational misalignment. To understand the effects of alignment errors on the accuracy of measured circumferences of the human lower body, such misalignment is simulated in this paper and the resulting characteristic error patterns are analyzed. The results show that the total error consists of two components, namely translational and tilt. Linear correlations were found between the translational error (R2 = 0.90, … 0.97) and the change in circumferences as well as between the tilt error (R2 = 0.55, … 0.78) and the change in the body's mean outline. Finally, by systematic analysis of the error patterns, recommendations were derived and applied to 3D body scans of human subjects resulting in a reduction of error by 67% and 84%.