Related Experiment Video
Updated: Jul 7, 2025

Author Spotlight: Implementation of BIVA for Analyzing Disease Risk Factors in Patients with Low Body Cell Mass
Published on: July 14, 2023
The Effect of a Compression Bandage on Dual Energy X-ray Absorptiometry Body Composition Scores
Yamileth Chacón-Araya1, Elizabeth Carpio-Rivera1, Andrea Quirós-Quirós2
1Human Movement Sciences Research Center (CIMOHU), University of Costa Rica, Costa Rica; School of Physical Education and Sports, University of Costa Rica, Costa Rica.
Background:
Dual-energy X-ray absorptiometry (DXA) measures are affected by the noise produced by external factors such as textile compression found in loose clothing. The study aimed to determine the effect of a compression bandage (CB) on body composition (BC) assessed by DXA.
Methods:
Sixty volunteers (age=21.4±4.7yr.) underwent full-body DXA scans on a control (CTRL) condition and after wearing a 30-mmHg CB on the trunk, legs, and arms. ANOVA (2 genders by 2 experimental conditions) determined mean interactions in BC variables tissue body fat% (BF%), region body fat% (RBF%), body tissue (BT), fat mass (FM), lean mass (LM), bone mineral content (BMC), and total mass (TM). Absolute reliability in BC scores was studied by the typical error of the measurement (TEM), the coefficient of variability (CV), and Bland-Altman plots.
Results:
ANOVA interactions were found on tissue total BF% (p=0.049), RBF% (p=0.048), android lean mass (p=0.004), and android total mass (p=0.019). The CV was small for tissue BF% (2.61±0.93%, CI95%=0.79, 4.43%), RBF% (2.66±1.78%, CI95%=-0.83, 6.15%), BT (4.82±2.19%, CI95%=0.54, 9.10kg), FM (4.17±2.25%, CI95%=-0.24, 8.58kg), LM (3.25±2.44%, CI95%=-1.53, 8.04kg), BMC (4.81±2.96%, CI95%=-0.99, 10.62kg), and TM (2.84±2.80%, CI95%=-2.65, 8.33kg). Bland-Altman plots showed random error for BF%, LM, and BMC. A bias of 0.5% was observed on BF% in males.
Conclusion:
A CB worn during a full-body DXA scan elicited similar BC scores than not wearing it. The variation in scores was <10% for most BC variables, and a trivial bias of 0.5% in BF% was detected in male scores.