Local Diagnostic Reference Levels for Pediatric Interventional Cardiology Procedures in Argentina

Patricia Azcurra1, Fernando Leyton2, Victorio Lucini1

  • 1Interventional Cardiology Department, Hospital Italiano de Buenos Aires, Buenos Aires C1199ABB, Argentina.

PubMed

Insights

This study establishes preliminary diagnostic reference levels (DRL) for pediatric interventional cardiology in Argentina. These radiation dose levels aim to minimize patient exposure during fluoroscopy-guided procedures.

Area of Science:

  • Medical Physics
  • Radiology
  • Pediatric Cardiology

Background:

  • Radiation dose optimization is crucial in pediatric interventional cardiology.
  • Establishing local diagnostic reference levels (DRL) aids in dose management.
  • The OPRIPALC program promotes radiation protection in Latin America.

Purpose of the Study:

  • To propose preliminary local diagnostic reference levels (DRL) for pediatric interventional cardiology (PIC) procedures in Argentina.
  • To stratify DRLs based on patient age and weight ranges.
  • To contribute to radiation dose optimization in pediatric fluoroscopy.

Main Methods:

  • Data collection within the OPRIPALC program framework.
  • Calculation of DRLs based on pediatric patient weight and age groups.
  • Comparison of local DRLs with international findings.

Main Results:

  • Preliminary DRLs were established for various pediatric weight and age groups.
  • Specific DRL values presented: 7.1 Gy·cm² (<5 kg), 10.7 Gy·cm² (5-15 kg), 18.0 Gy·cm² (15-30 kg), 15.9 Gy·cm² (30-50 kg), 28.2 Gy·cm² (50-80 kg).
  • Age-based DRLs: 5.3 Gy·cm² (<1 yr), 11.2 Gy·cm² (1-5 yrs), 19.6 Gy·cm² (5-10 yrs), 21.4 Gy·cm² (10-16 yrs).

Conclusions:

  • The study provides initial DRLs for PIC in Argentina.
  • The established DRLs are comparable to international values.
  • Significant potential exists for further radiation dose optimization in pediatric interventional procedures.