Related Experiment Video
Updated: Jul 19, 2026

Voluntary Breath-hold Technique for Reducing Heart Dose in Left Breast Radiotherapy
Published on: July 3, 2014
Investigation on the accuracy of field widths and the quantitative relationship with energy variations for dynamic
Ho Ming Poon1, Tin Lok Chiu1, Siu Ki Yu1
1Medical Physics Department, Hong Kong Sanatorium & Hospital, Hong Kong SAR, China.
Background And Purpose:
TomoEDGE is an advanced technology for TomoTherapy treatment delivery by introducing a sliding-window dynamic jaw motion. The front and back jaws move independently at the start and end of a target volume along the longitudinal couch direction to reduce the undesired dose to the normal tissues. The accuracy of field width is essential to treatment delivery in this regard. The purpose of this work was to analyze the performance of dynamic jaws on helical tomotherapy and investigate the relationship with energy variation.
Methods:
The Tomotherapy-Quality-Assurance (TQA) Dynamic Field Width procedure was performed monthly across three tomotherapy machines. All field widths were analyzed, especially the FWHM of the 10 mm field width. Field width measurements were compared with the ratio of Percentage Depth Dose at 20 and 10 cm to render the value of correlation. Changes in beam FWHM and energy were further discussed. Two-year data were collected for this purpose.
Results:
On average, measured field widths in each unit agreed within 1% tolerance recommendation stated. The average absolute difference between reference and measured FWs in each unit was approximately 0.07 mm. An increase of 1.5% in the FW of the 10 mm nominal beam width was correlated with a 1% increase in PDD20,10 ratio, implying a positive correlation between the two factors (p < 0.002).
Conclusions:
A positive correlation between nominal 10 mm FW and PDD20,10 was observed. In the case that the PDD20,10 marginally passes the QA tests, users are recommended to consider further verification on Dynamic Jaws to ensure the smallest field width to be within tolerance, which is essential to maintain effective treatment in TomoEDGE system. Since the regression of this study was a single-factor model, other confounding factors such as the focal spot size of linear accelerator should also be considered when evaluating the machine status.
More Related Videos
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation I: X-ray and CT
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Imaging Studies III: Computed Tomography

