Left bundle branch pacing lead for sensing ventricular arrhythmias in implantable cardioverter-defibrillator: A pilot
Shunmuga Sundaram Ponnusamy1, Vadivelu Ramalingam1, Selvaganesh Mariappan1
1Department of Cardiology, Velammal Medical College Hospital and Research Institute, Madurai, Tamil Nadu, India.
Background:
Left bundle branch pacing (LBBP) has been suggested as an alternative modality for biventricular pacing in cardiac resynchronization therapy (CRT)-eligible patients. As it provides stable R-wave sensing, LBBP has been recently used to provide sensing of ventricular arrhythmia in patients receiving implantable cardioverter-defibrillator (ICD) with CRT.
Objective:
The aim of this study was to analyze the long-term safety and efficacy of the LBBP lead for appropriate detection of ventricular arrhythmia and delivery of antitachycardia pacing (ATP) in patients requiring defibrillator therapy with CRT.
Methods:
CRT-eligible patients who underwent successful LBBP-optimized ICD and LBBP-optimized CRT with defibrillator were enrolled. The LBBP lead was connected to the right ventricular-P/S port after capping the IS-1 connector plug of the DF-1-ICD lead. LBBP-optimized ICD or LBBP-optimized CRT with defibrillator was decided on the basis of correction of conduction system disease. Documented arrhythmic episodes and therapy delivered were analyzed.
Results:
Thirty patients were enrolled. The mean age was 59.7 ± 10.5 years. LBBP resulted in an increase in left ventricular ejection fraction from 29.9% ± 4.6% to 43.9% ± 11.2% (P < .0001). During a mean follow-up of 22.9 ± 12.5 months, 254 ventricular arrhythmic events were documented. Appropriate events (n = 225 [89%]) included nonsustained ventricular tachycardia (VT) (n = 212 episodes [94%]), VT (n = 8 [3.5%]), and ventricular fibrillation (n = 5 [2.5%]). ATP efficacy in terminating VT was 75%. Eleven percent of episodes (n = 29) were inappropriately detected because of T-wave oversensing. Inappropriate therapy (ATP) was delivered for 14 episodes (5.5%). Three patients (10%) had worsening of tricuspid regurgitation.
Conclusion:
Sensing from the LBBP lead for arrhythmia detection is safe as ∼90% of the episodes were detected appropriately. Future studies with a dedicated LBBP-defibrillator lead along with algorithms to avoid oversensing can help in combining defibrillation with conduction system pacing.
More Related Videos
16:40A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
10:05Testing the Efficacy of Pharmacological Agents in a Pericardial Target Delivery Model in the Swine
Published on: July 7, 2016
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin...
Electrophysiology of Normal Cardiac Rhythm
Conduction System of the Heart
This system relies on the unique properties of nodal and Purkinje cells:...
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow...
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...