Background:
Neoatherosclerosis, a prominent contributor to in-stent restenosis (ISR), persists as a formidable challenge during percutaneous coronary intervention. Excimer laser coronary atherectomy (ELCA) and embolic protection devices may help reduce coronary flow disturbance from procedure-related distal embolization.
Case Summary:
A 71-year-old man experienced in-stent neoatherosclerosis rupture-related non-ST segment elevation myocardial infarction. Multidisciplinary intracoronary imaging, including intravascular ultrasound and optical coherence tomography (OCT), suggested that the ISR was caused by a neoatherosclerosis rupture that can potentially lead to distal embolization. Excimer laser coronary atherectomy (fluence, 45 mJ/mm2 and frequency, 25 pulse/s) using a 1.7 mm concentric catheter was performed with distal protection using Filtrap (Nipro Corporation, Tokyo, Japan), which significantly reduced the volume of the neoatherosclerosis. However, subsequent ELCA on the highest setting (fluence, 60 mJ/mm2 and frequency, 40 pulse/s) led to a filter no-reflow phenomenon, although OCT revealed a further effective vaporization of the neoatherosclerosis and an apparent reduction of soft tissue compatible with the thrombus. After removing the embolic protection device, drug-coated balloon angioplasty provided optimal results without coronary flow disturbance.
Discussion:
Excimer laser coronary atherectomy reduces soft plaque and thrombus burden, which can reduce the occurrence of distal embolization in select cases. In the case of this patient, procedure-related distal embolization may have been induced by the heightened photomechanical effects resulting from the use of the highest setting in ELCA under increased intracoronary arterial pressure caused by continuous saline injection during ELCA. Concomitant distal protection during ELCA may be more feasible for preventing coronary flow disturbance in patients with a large amount of neoatherosclerosis.
More Related Videos
13:48Reduction of Radiation Exposure during Endovascular Treatment of Peripheral Arterial Disease Combining Fiber Optic RealShape Technology and Intravascular Ultrasound
Published on: April 21, 2023
06:26Novel and Innovative Hybrid Technique for Type A Aortic Dissection
Published on: March 28, 2025
Related Concept Videos
Coronary Artery Disease V: Interprofessional Care
Atherosclerosis III: Management
Peripheral Artery Disease III: Interprofessional Care