A New Portable In Vitro Exposure Cassette for Aerosol Sampling
Published on: February 22, 2019
Aerosol and surface contamination assessment of a novel ventilated infectious aerosol containment device
Runcheng Fang1, Niles Andrus1, Thomas Dominguez1
1Division of Occupational & Environmental Health, Department of Family & Preventive Medicine, University of Utah, Salt Lake City, UT 84108, USA.
Abstract:
Healthcare personnels (HCPs) are at risk of respiratory infectious diseases during patient care activities. HCPs rely primarily on personal protective equipment to prevent pathogen exposures, but there is a need to develop alternative, or complementary control strategies, including engineering controls. The objective of this study was to evaluate the ability of the 3 designs (denoted D1A, D1B, and D2) of the University of Utah Containment Ventilation for Exposure Reduction (U-COVER), a protective barrier enclosure device to contain respirable aerosols when placed over a simulated patient. The 2 primary performance metrics were the percent reduction in: (i) the concentration of respirable aerosols in the simulated breathing zone of an HCP, and (ii) surface contamination outside the device, which were tested using salt aerosols and fluorescein aerosols, respectively. Briefly, salt or fluorescein aerosols were generated as though expelled by a prone patient under 3 conditions: (i) no device (control), (ii) with the device but without exhaust ventilation, and (iii) with the device with exhaust ventilation. Device D2 was also tested under simulated use conditions, in which cardboard "arms" were placed inside the device ports. All 3 device designs showed the ability to reduce particle concentrations in the simulated HCP breathing zone and on surfaces by >99% with exhaust ventilation compared to the control condition. Without exhaust ventilation, device performance was lower and highly variable. Under simulated use conditions, device D2 reduced particle concentrations in the simulated HCP breathing zone by ≥91% and on surfaces by >99% relative to control for all combinations of "arms" tested. The U-COVER device demonstrates excellent aerosol containment and warrants further testing with dynamic simulated or actual use conditions.
Related Concept Videos
Transmission-based Precautions II: Airborne and Protective Environment
Airborne precautions:
Use airborne precautions when treating patients known or suspected to have diseases that spread through the air—for example, tuberculosis or measles. These organisms are present in smaller droplets expelled by an infected person and...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...