Related Experiment Video
Updated: Jul 6, 2025

Laparoscopic-Assisted Seldinger Technique for Peritoneal Dialysis Catheter Insertion
Published on: May 23, 2025
[Catheter Replacement Methods in Hemodialysis Patients With Dysfunctional Tunneled-Cuffed Catheters With Fibrin
Lizhu Jin1,2, Hui Wang1, Tianlei Cui1
1( 610041) Department of Nephrology, West China Hospital, Sichuan University, Chengdu 610041, China.
Insights
Replacing tunneled-cuffed catheters (TCCs) via a fibrin sheath crevice with balloon dilation offers superior long-term blood flow for hemodialysis. This method is safe and effective, outperforming in situ replacement in maintaining optimal catheter function.
Area of Science:
- Nephrology
- Vascular Surgery
- Medical Devices
Objective:
Tunneled-cuffed catheters (TCCs) are frequently used for establishing hemodialysis access for maintenance hemodialysis in older patients with exhausted resources of peripheral vessels. Fibrin sheath formation around the catheter is one of the most common complications of long-term use of indwelling catheter, which may cause the malfunction of the catheter. In this study, we intend to compare the prognosis of two catheter replacement methods, in situ replacement and replacement through a fibrin sheath crevice, with both being assisted by balloon dilation, and to explore the optimal catheter replacement process.
Methods:
A retrospective study was conducted with 52 patients who underwent a replacement of their TCCs. Among them, 27 cases had their TCC replaced by the modified method of replacement through a fibrin sheath crevice and were referred to as the sheath crevice group, while 25 cases underwent in situ catheter replacement and were referred to as the in situ group. The primary outcome indicators included maximum blood flow in hemodialysis catheter and the urea clearance rate calculated by Kt/V values at the 1, 3, and 6-month follow-ups. The secondary outcomes included dialyzer alarms being set off and catheter-related infections during follow-up.
Results:
There was no significant difference between the general data of the two groups. There was no massive blood loss during the replacement procedure. Neither were there cardiac tamponade, catheter-associated infections, or other complications. Follow-ups were made 1, 3, and 6 months after the replacement procedure. The sheath crevice group had higher catheter blood flow and Kt/V values at the 6-month follow-up than the in situ group did ([241.85±9.62] mL/min vs. [234.40±11.21] mL/min, P=0.014 and 1.31±0.55 vs. 1.27±0.49, P=0.005, respectively). During the follow-up process, access alarms were reported in 5 patients (three in the in situ group and two in the sheath crevice group) during dialysis. No catheter-associated infection occurred in either group.
Conclusion:
The catheter replacement method of balloon dilation-assisted catheter insertion through a fibrin sheath crevice is safe and effective, resulting in better long-term catheter blood flow compared with that of in situ catheter replacement.
Related Concept Videos
Dialysis
Acute kidney injury develops suddenly and can be caused by pre-renal causes (e.g., hypovolemia, shock), intrinsic renal causes (e.g., acute tubular necrosis), or post-renal causes (e.g., urinary obstruction). In contrast, chronic renal failure progresses gradually over time and is often...
Parentral Nutrition: Centeral and Peripheral Parental Nutrition
PN can be administered through two primary routes:
1. Central Parenteral Nutrition (CPN):
CPN involves delivering a high concentration of nutrients through a large vein. This is typically achieved using a Peripherally Inserted Central Catheter (PICC) or,...