Related Experiment Video
Updated: Jul 6, 2025

05:36
Standardized Model of Ventricular Fibrillation and Advanced Cardiac Life Support in Swine
Published on: January 30, 2020
7.8K
Comparative usability of manual defibrillators - A human factors study
Stefan Schumann1, Axel Schmutz1, Kim Feger1
1Department of Anesthesiology and Critical Care, Medical Center - University of Freiburg, Faculty of Medicine, University of Freiburg, Germany.
Resuscitation Plus
|January 1, 2024
Summary
Manual defibrillator usability varies significantly. The Physio-Control Lifepak 20e demonstrated superior ease of use and faster operation compared to other devices, highlighting the impact of interface design on critical care efficiency.
Area of Science:
- Medical Devices
- Human-Computer Interaction
- Emergency Medicine
Background:
- Manual defibrillators are critical for cardiac arrest resuscitation.
- Improper or slow defibrillator operation can negatively impact patient care.
- Self-explanatory interfaces are crucial for reducing operational errors.
Purpose of the Study:
- To evaluate and compare the usability of four commercially available manual defibrillators.
- To identify how interface design affects operator performance and efficiency.
- To determine which defibrillator offers the best user experience in simulated emergency scenarios.
Main Methods:
- 31 medical students performed 15 tasks on four defibrillators (Lifepak 20e, Defigard Touch 7, Corpuls 3, X-Series).
- Eye-tracking monitored operator gaze, while task completion times and assistance frequency were recorded.
- Subjective usability was assessed using a standardized questionnaire.
Main Results:
- The Lifepak 20e required the fewest assistances and had the shortest task completion times (124 ± 31 s).
- The Zoll X-Series required the most assistances (63) and had the longest completion times (289 ± 85 s).
- Eye-tracking identified interface issues impacting performance; subjective usability ratings correlated with objective measures.
Conclusions:
- Defibrillator usability varies significantly, with task-specific differences observed.
- Interface design directly impacts operator efficiency and perceived usability.
- Addressing usability issues in defibrillator interfaces can enhance operator performance and potentially improve patient outcomes.