Background:
The aim of this study was to validate the use of temperature sensors to accurately measure thermoplastic volar forearm splint wear in a healthy cohort of volunteers using 5- and 15-minute temperature measurement intervals.
Methods:
A prospective diagnostic study was performed to evaluate the diagnostic accuracy of temperature sensors in monitoring splint wear in 8 healthy volunteers between December 2022 and June 2023. Temperature sensors were molded into thermoplastic volar forearm splints. Volunteers who were familiar with the study aims were asked to keep an exact log of the time spent wearing the splint ("actual wear time"). Sensors recorded temperatures every 5 or 15 minutes, and separate algorithms were developed to determine the sensor-detected wear time compared with the actual wear time as the gold standard. The algorithms were then externally validated with the total population.
Results:
The 5-minute and 15-minute algorithms demonstrated excellent sensitivity (99.1% vs 96.6%), specificity (99.9% vs 99.9%), positive (99.4% vs 99.5%) and negative (99.9% vs 99.3%) predictive value, and diagnostic accuracy (99.8% vs 99.3%), respectively. The 5-minute algorithm recorded 99.5% of the total splint hours, whereas the 15-minute algorithm recorded 96.1%. There was no significant difference between the actual time per wear session (5.4 ± 2.7 hours) and the time estimated by the 5-minute algorithm (5.4 ± 2.6 hours; P = .40), but there was a significant difference for the 15-minute algorithm (5.2 ± 2.6 hours; P < .001).
Conclusion:
Temperature sensors can be used to accurately monitor thermoplastic volar forearm splint wear.
Level Of Evidence:
Diagnostic II.
Related Concept Videos
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's...
Assessing Body Temperature - Axilla
Step 1: Perform hand hygiene and put on clean gloves to maintain infection control and prevent cross-contamination.
Step 2: Prepare the patient by explaining the procedure to ensure understanding and cooperation. Ensure privacy, expose the axilla, and inform the patient that minimal movement is crucial for an accurate reading.
Step 3: Adjust the patient’s clothing to expose only the axilla. It minimizes...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...