Related Experiment Video
Updated: Jul 6, 2025

Author Spotlight: Standardization and Best Practices for Advancing Lung Imaging Using 129Xe MRI
Published on: November 21, 2023
Major pneumothorax during pediatric cardiac MRI procedure under general anesthesia: step-by-step analysis and
Quentin Delhez1, Laurent Bairy2, John Mitchell2
1Department of Anesthesiology, Université Catholique de Louvain, Centre Hospitalier Universitaire UCL Namur site Godinne, Yvoir, Belgium. quentin.delhez@chuuclnamur.uclouvain.be.
Background:
To perform step-by-step analysis of the different factors (material, anesthesia technique, human, and location) that led to major pneumothorax during an infrequent pediatric cardiac MRI and to prevent its occurrence in the future. Anesthesia equipment used in a remote location is often different than those in operating rooms. For magnetic resonance imaging (MRI), ventilation devices and monitors must be compatible with the magnetic fields. During cardiac MRI numerous apneas are required and, visual contact with the patient is limited for clinical evaluation. Anesthesia-related barotrauma and pneumothorax are rare in children and the first symptoms can be masked.
Case Presentation:
A 3-year-old boy with atrial septal defect (ASD) and suspicious partial anomalous pulmonary venous return was anesthetized and intubated to perform a follow up with MRI. Sevoflurane maintenance and ventilation were performed using a circular CO2 absorber device, co-axial circuit, and 500 mL pediatric silicone balloon. Apneas were facilitated by Alfentanyl boluses and hyperventilation. A few moderated desaturations occurred during the imaging sequences without hemodynamic changes. At the end of the MRI, facial subcutaneous emphysema was observed by swollen eyelids and crackling snow neck palpation. A complete left pneumothorax was diagnosed by auscultation, sonography examination, and chest radiograph. Pneumo-mediastinum, -pericardium and -peritoneum were present. A chest drain was placed, and the child was extubated and transferred to the pediatric intensive care unit (PICU). Despite the anesthesiologist's belief that PEEP was minimal, critical analysis revealed that PEEP was maintained at a high level throughout anesthesia. After the initial barotrauma, repeated exposure to high pressure led to the diffusion of air from the pleura to subcutaneous tissues and mediastinal and peritoneal cavities. Equipment check revealed a functional circular circuit; however, the plastic adjustable pressure-limiting valve (APL) closed within the last 30° rotation. The balloon was found to be more rigid and demonstrated significantly reduced compliance.
Conclusions:
Anesthetists require proficiency is using equipment in non-OR locations and this equipment must be properly maintained and checked for malfunctions. Controlling the human factor risks by implementing checklists, formations, and alarms allows us to reduce errors. The number of pediatric anesthesia performed routinely appeared to be essential for limiting risks and reporting our mistakes will be a benefit for all who care about patients.
Insights
High PEEP and equipment malfunction caused pneumothorax during pediatric cardiac MRI. Careful equipment checks and anesthesia protocols are crucial for patient safety in non-OR settings.
Area of Science:
- Pediatric Anesthesiology
- Cardiovascular Imaging
- Medical Device Safety
Background:
- Pediatric cardiac MRI requires specialized anesthesia techniques due to magnetic field compatibility and limited patient monitoring.
- Anesthesia-related barotrauma and pneumothorax are rare but serious complications in children, with symptoms that can be masked.
- Remote anesthesia locations may use different equipment than operating rooms, necessitating thorough checks.
Observation:
- A 3-year-old boy undergoing cardiac MRI developed facial subcutaneous emphysema, diagnosed as left pneumothorax, pneumo-mediastinum, -pericardium, and -peritoneum.
- Ventilation during the MRI used a circular CO2 absorber, co-axial circuit, and pediatric silicone balloon, with Alfentanyl for apnea facilitation.
- Despite perceived minimal PEEP, analysis revealed high levels, and equipment checks identified a malfunctioning adjustable pressure-limiting valve and a rigid, less compliant balloon.
Findings:
- High positive end-expiratory pressure (PEEP) and equipment malfunction, specifically a faulty adjustable pressure-limiting valve (APL) and a stiff anesthesia bag, contributed to barotrauma.
- Repeated high-pressure ventilation led to air diffusion from the pleura into subcutaneous tissues and body cavities.
- The combination of anesthesia technique, equipment issues, and the unique demands of cardiac MRI created a risk for pneumothorax.
Implications:
- Anesthetists must be proficient with equipment in non-operating room settings and perform rigorous pre-procedure checks.
- Implementing checklists, standardized training, and alarms can mitigate human factor risks in pediatric anesthesia.
- Routine pediatric anesthesia experience and transparent reporting of errors are vital for improving patient safety during complex procedures like cardiac MRI.
More Related Videos
Related Concept Videos
Pneumothorax-I
Pneumothorax can be even further classified as spontaneous, traumatic, and tension pneumothorax.
Pneumothorax-II
Clinical Manifestations:
Endoscopic Studies II: Thoracocentesis
Description
Excess pleural fluid or air may accumulate in some respiratory disorders in the thoracic cavity. To treat pleural effusion, a physician conducts thoracentesis by carefully piercing the chest wall and entering...
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.