Cone-beam x-ray luminescence computed tomography (CB-XLCT) prototype development and performance evaluation.
Yu-Hong Wang1,2, David Shih-Chun Jin1,3, Tian-Yu Wu4
1Department of Biomedical Imaging and Radiological Sciences, National Yang Ming Chiao Tung University, 112304 Taipei, Taiwan, ROC.
A new rotational cone-beam x-ray luminescence computed tomography (CB-XLCT) system prototype was developed for pre-clinical theranostic imaging. This system shows potential for high-quality imaging in small animals with acceptable radiation doses.
Area of Science:
- Medical Imaging
- Biomedical Engineering
- Radiological Physics
Background:
- X-ray luminescence computed tomography (XLCT) offers potential for theranostic applications.
- Developing integrated XLCT systems requires addressing imaging chain calibration and optimal imaging conditions.
- Pre-clinical studies are crucial for advancing XLCT technology towards clinical translation.
Purpose of the Study:
- To develop and evaluate a prototype rotational cone-beam x-ray luminescence computed tomography (CB-XLCT) system.
- To establish a geometric calibration method for both XL and CT imaging modalities.
- To assess the system's performance and commercialization feasibility for pre-clinical theranostic imaging.
Main Methods:
- A rotational CB-XLCT system prototype was designed using a low-dose x-ray tube and a rotating gantry.
- Monte Carlo GATE simulations determined optimal imaging conditions for nanophosphor particles (NPs).
- Image cross-correlation (ICC) was employed for automatic geometric calibration of XL and CT images.
Main Results:
- The system achieved high linearity (0.99) and a contrast-to-noise ratio (CNR) of 19.5 in phantom experiments.
- XL images demonstrated a Dice similarity coefficient (DICE) of 0.149 for 1 mm separation between light sources.
- Demonstrated XLCT imaging results using Letter phantoms with NPs.
Conclusions:
- The developed CB-XLCT prototype shows promise for high-quality pre-clinical imaging in small animals.
- CT image performance is comparable to commercial systems; XL images show potential but require further development for biomedical applications.
- The system's feasibility for commercialization was assessed, highlighting areas for future research.