[Accuracy evaluation of static guided implant placement using an intraoral scan method]
Sheng-Jia Ye1, Xiao-Yan Yang, Jia-Ming Zhang
1Department of Prosthodontics, Shanghai Ninth People's Hospital, Shanghai Jiao Tong University School of Medicine; College of Stomatology, Shanghai Jiao Tong University; National Center for Stomatology; National Clinical Research Center for Oral Diseases; Shanghai Key Laboratory of Stomatology; Shanghai Research Institute of Stomatology. Shanghai 200011, China.
Purpose:
To evaluate the accuracy of static guided implant placement with intraoral scanning technology and to analyze the influencing factors of guided surgery.
Methods:
Totally 27 cases were included in this retrospective study. The implant designs were made in 3Shape Implant Studio and then guided implant surgeries were performed with CAD-CAM templates. Postoperative implant positions were detected with an intraoral scanner (3Shape TRIOS) and deviation of implantation was evaluated using established CAD/CAM based evaluation method. SAS 9.4 software package was used for data analysis.
Results:
The mean deviation of entrance point and apical point was (1.182±0.609) mm and (1.658±0.741) mm, respectively. Angular deviation was (5.712±3.347)°. Implant quadrant, location of the implant site, guidance degree, supporting type and implant size influenced direction deviation, while angular deviation was mainly affected by guidance degree and number of missing teeth.
Conclusions:
Accuracy of static guided implant placement can be influenced by many factors. More research is needed to improve the accuracy of static guided implantation.