Patient-specific automated cerebrospinal fluid pressure control to augment spinal wound closure: a case series using
Danyal Z Khan1,2, Kanza Tariq1,3, Keng Siang Lee4
1Department of Neurosurgery, National Hospital for Neurology and Neurosurgery, London, UK.
Objective:
Spinal cerebrospinal fluid (CSF) leaks are common, and their management is heterogeneous. For high-flow leaks, numerous studies advocate for primary dural repair and CSF diversion. The LiquoGuard7® allows automated and precise pressure and volume control, and calculation of patient-specific CSF production rate (prCSF), which is hypothesized to be increased in the context of durotomies and CSF leaks.
Methods:
This single-centre illustrative case series included patients undergoing complex spinal surgery where: 1) a high flow intra-operative and/or post-operative CSF leak was expected and 2) lumbar CSF drainage was performed using a LiquoGuard7®. CSF diversion was tailored to prCSF for each patient, combined with layered spinal wound closure.
Results:
Three patients were included, with a variety of pathologies: T7/T8 disc prolapse, T8-T9 meningioma, and T4-T5 metastatic spinal cord compression. The first two patients underwent CSF diversion to prevent post-op CSF leak, whilst the third required this in response to post-op CSF leak. CSF hyperproduction was evident in all cases (mean >/=140ml/hr). With patient-specific CSF diversion regimes, no cases required further intervention for CSF fistulae repair (including for pleural CSF effusion), wound breakdown or infection.
Conclusions:
Patient-specific cerebrospinal fluid drainage may be a useful tool in the management of high-flow intra-operative and post-operative CSF leaks during complex spinal surgery. These systems may reduce post-operative CSF leakage from the wound or into adjacent body cavities. Further larger studies are needed to evaluate the comparative benefits and cost-effectiveness of this approach.