Iodine contrast volume reduction in preoperative transcatheter aortic valve implantation computed tomography:
T Yoshiura1, T Masuda2, Y Kobayashi3
1Graduate School of Health Sciences, Kumamoto University, Kuhonji 4-24-1, Chuo-ku, Kumamoto, 860-0976, Japan; Department of Medical Technology, Tsuchiya General Hospital, Nakajima-cho 3-30, Naka-ku, Hiroshima, 730-8655, Japan.
Introduction:
This study aimed to compare the vascular enhancement and radiation dose in preoperative transcatheter aortic valve implantation (TAVI) computed tomography (CT) with a reduced contrast medium (CM) using volume scans in 256-multidetector row CT (MDCT) with a standard CM using 64-MDCT.
Methods:
This study included 78 patients with preoperative TAVI CT with either 64- or 256-MDCT. The CM was injected at 1.5 mL/kg in the 64-MDCT group and 1.0 mL/kg in the 256-MDCT group. We compared vascular enhancement of the aortic root and access routes, image quality (IQ) scores, and radiation dose in both groups.
Results:
Despite the reduced CM (by 33 %) in the 256-MDCT group, the mean vascular enhancement of the right and left subclavian arteries was significantly higher than that in the 64-MDCT group [284 and 267 Hounsfield units (HU) vs. 376 and 359 HU; p < 0.05]; however, no significant differences in the mean vascular enhancement in the ascending aorta, abdominal aorta at the celiac level, and bilateral common femoral arteries were observed between the two groups (p > 0.05 for all). The median IQ scores at the aortic root were higher in the 256-MDCT group than in the 64-MDCT group (3 vs. 4; p < 0.05), and those at the femoral access routes were comparable (4 vs. 4; p = 0.33). The mean effective dose was significantly reduced by 30 % in the 256-MDCT group (23.6 vs. 16.3 mSv; p < 0.05).
Conclusion:
In preoperative TAVI CT, volume scans using 256-MDCT provide comparable or better vascular enhancement and IQ with a 30 % reduction in CM and radiation dose than those using 64-MDCT.
Implications For Practice:
Volume scan using 256-MDCT for preoperative TAVI CT may reduce CM and radiation dose in TAVI patients with renal dysfunction.
More Related Videos
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
06:59Improved Registration of 3D CT Angiography with X-ray Fluoroscopy for Image Fusion During Transcatheter Aortic Valve Implantation
Published on: June 3, 2018
Related Concept Videos
Radiological Investigation I: X-ray and CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...