Copper Isotope Compositions Measured Using a Sapphire Dual Path MC-ICPMS with a Collision/Reaction Cell
Tu-Han Luu1, Daniel Peters2, Esther Lahoud1
1Université Paris Cité, Institut de Physique du Globe de Paris, CNRS, F-75005 Paris, France.
Abstract:
We present a new approach to Cu isotopic measurements using a state-of-the-art Nu Sapphire multicollector inductively coupled plasma source mass spectrometer equipped with a collision/reaction cell (CRC-MC-ICPMS). We investigate the effects of Na doping and Cu concentration mismatch between bracketing standard and unknown samples and demonstrate the efficacy of introducing a He-H2 gas mix into the CRC to efficiently eliminate the sample matrix-based 40Ar23Na+ isobaric interference on 63Cu+. This capability is crucial when measuring samples with high Na/Cu ratios, such as some biological samples, which have significantly different chemical compositions compared to most geological samples. Moreover, considering the necessity of obtaining large data sets for biological samples to ensure reliable interpretations, the implementation of a CRC for mitigating the 40Ar23Na+ interference offers the advantage of minimizing the requirement for extensive Cu chemical separation procedure prior to Cu isotopic measurements. Our results demonstrate that the accurate determination of the δ65Cu values is achievable for samples with Na/Cu concentration ratios of up to ∼65, even when measuring 100 ppb Cu solutions (equivalent to a signal of ∼3.5-4 V total Cu). Furthermore, our results showcase a good short-term repeatability on δ65Cu for pure Cu standard solutions (NIST SRM 976 and Cu-IPGP), typically of 0.05‰ (2 SD) when measuring >50 ppb Cu solutions. Our long-term external reproducibility stands at approximately 0.07‰ (2 SD). This value accounts for the variable Cu concentrations analyzed across the different analytical sequences (from 10 to 100 ppb Cu solutions). To validate the robustness of our analytical method, we first conduct a comparison between data sets from mice brains processed twice through column chemistry using a Thermo Finnigan Neptune MC-ICPMS and a Nu Sapphire CRC-MC-ICPMS in CRC mode. This comparison serves to verify the reliability of our method for measuring Cu isotopic composition using the CRC on samples with a low Na/Cu ratio after traditional chemical processing. Then, we compare the data sets obtained for biological standards (tuna fish ERM-CE 464 (IRMM) and human serum Seronorm Trace Elements Serum L-1) processed either once, or twice, through column chemistry and demonstrate that the CRC allows accurate Cu isotopic measurements of the samples processed only once and therefore with a higher Na/Cu ratio.
Related Concept Videos
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
Atomic Emission Spectroscopy: Lab
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....
Tandem Mass Spectrometry
Secondary fragmentations occur in the interaction cell and can be induced by various factors. Fragmentation induced by collision with inert gases, such as N2, Ar, He, etc., is called collision-induced...