Assessing the effect of scanning parameter on the size and density of pulmonary nodules: a phantom study
Donghua Meng1, Zhen Wang2, Changsen Bai3
1Department of Radiology, Tianjin Medical University Cancer Institute and Hospital, National Clinical Research Center for Cancer, Key Laboratory of Cancer Prevention and Therapy, Tianjin's Clinical Research Center for Cancer, Tianjin, China.
Background:
Lung cancer remains a leading cause of death among cancer patients. Computed tomography (CT) plays a key role in lung cancer screening. Previous studies have not adequately quantified the effect of scanning protocols on the detected tumor size. The aim of this study was to assess the effect of various CT scanning parameters on tumor size and densitometry based on a phantom study and to investigate the optimal energy and mA image quality for screening assessment.
Methods:
We proposed a new model using the LUNGMAN N1 phantom multipurpose anthropomorphic chest phantom (diameters: 8, 10, and 12 mm; CT values: - 100, - 630, and - 800 HU) to evaluate the influence of changes in tube voltage and tube current on the size and density of pulmonary nodules. In the LUNGMAN N1 model, three types of simulated lung nodules representing solid tumors of different sizes were used. The signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were used to evaluate the image quality of each scanning combination. The consistency between the calculated results based on segmentation from two physicists was evaluated using the interclass correlation coefficient (ICC).
Results:
In terms of nodule size, the longest diameters of ground-glass nodules (GGNs) were closest to the ground truth on the images measured at 100 kVp tube voltage, and the longest diameters of solid nodules were closest to the ground truth on the images measured at 80 kVp tube voltage. In respect to density, the CT values of GGNs and solid nodules were closest to the ground truth when measured at 80 kVp and 100 kVp tube voltage, respectively. The overall agreement demonstrates that the measurements were consistent between the two physicists.
Conclusions:
Our proposed model demonstrated that a combination of 80 kVp and 140 mA scans was preferred for measuring the size of the solid nodules, and a combination of 100 kVp and 100 mA scans was preferred for measuring the size of the GGNs when performing lung cancer screening. The CT values at 80 kVp and 100 kVp were preferred for the measurement of GGNs and solid nodules, respectively, which were closest to the true CT values of the nodules. Therefore, the combination of scanning parameters should be selected for different types of nodules to obtain more accurate nodal data.