Is the Blood Pressure-Enabled Smartwatch Ready to Drive Precision Medicine? Supporting Findings From a Validation
Wan Ling Lee1, Mahmoud Danaee2, Adina Abdullah3
1Department of Nursing Science, Faculty of Medicine, Universiti Malaya, 50603 Kuala Lumpur, Malaysia.
Background:
The popular wrist-worn wearables recording a variety of health metrics such as blood pressure (BP) in real time could play a potential role to advance precision medicine, but these devices are often insufficiently validated for their performance to enhance confidence in its use across diverse populations. The accuracy of BP-enabled smartwatch is assessed among the multi-ethnic Malaysians, and findings is discussed in comparison with conventional automated upper-arm BP device.
Methods:
Validation procedures followed the guidelines by the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Universal Standard (ISO 81060-2:2018). Quota sampling was employed to obtain eligible patients with normal and abnormal BP as per guideline. The measurements of BP were taken at wrist using HUAWEI WATCH D (test BP); and the readings were assessed against reference BP by the mercury sphygmomanometer. Agreement statistics and linear regression analyses were performed.
Results:
BP measurements (234 data pairs) from 78 patients that fulfilled AAMI/ESH/ISO protocol were analyzed. The BP readings taken by the HUAWEI WATCH D were comparable to reference BP by sphygmomanometer based on 1) Criterion 1: systolic blood pressure (SBP) = -0.034 (SD 5.24) and diastolic blood pressure (DBP) = -0.65 (SD 4.66) mm Hg; and 2) Criterion 2: SBPs = -0.034 (SD 4.18) and DBPs = -0.65 (SD 3.94) mm Hg. Factors of sociodemographic characteristics, anthropometric measurements, cardiovascular comorbidities, and wrist hair density were not significantly associated with the mean BP differences.
Conclusions:
HUAWEI WATCH D fulfilled criteria 1 and 2 of the AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018) guidelines. It can be recommended for clinical use across a wider population. The rich data from real-time BP measurements in concurrent with other health-related parameters recorded by the smartwatch wearable offer opportunities to drive precision medicine in tackling therapeutic inertia by personalizing BP control regimen.
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Pre-Procedural Guidelines for Assessing Blood Pressure
Measurement of Blood Pressure
Assessment of blood pressure in brachial artery(two-step method)