Preparation of Food Samples Using Homogenization and Microwave-Assisted Wet Acid Digestion for Multi-Element
Matjaž Rantaša1, David Majer1, Matjaž Finšgar2
1Faculty of Chemistry and Chemical Engineering, University of Maribor.
Abstract:
Sample preparation is crucial for elemental determination, and various techniques are available, one of which involves homogenization followed by acid digestion. Special care is required during sample handling in the preparation stage to eliminate or minimize potential contamination and analyte loss. Homogenization is a process that simultaneously reduces particle size and uniformly distributes sample components. Following homogenization, the sample undergoes acid digestion, wherein it is digested with acids and auxiliary chemicals at elevated temperatures, transforming solid samples into a liquid state. In this process, metals in the original sample react with acids to form water-soluble salts. Samples prepared through acid digestion are suitable for elemental analysis using techniques such as inductively coupled plasma mass spectrometry, inductively coupled plasma optical emission spectroscopy, atomic absorption spectroscopy, electrochemical methods, and other analytical techniques. This work details the preparation of food samples for multi-element determination using inductively coupled plasma mass spectrometry. The step-by-step procedure involves the homogenization process using a laboratory-sized mixer with ceramic blades, followed by acid digestion in closed vessels using microwave-assisted wet acid digestion. A mixture of 5.0 mL of 68 wt% HNO3 and 1.0 mL of 30 wt% H2O2 serves as an auxiliary reagent. This guide provides an explanation of the processes involved in both stages.
Related Concept Videos
Sample Preparation for Analysis: Advanced Techniques
Acid digestion with strong acids is commonly used to dissolve inorganic materials that are insoluble (do not dissolve) in water. This method can be useful for...
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Sample Preparation for Analysis: Overview
Bulk or large solid samples are typically reduced in size using grinding, crushing, or milling techniques to increase the...
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...